Supplementary Documentation

Page 44 of 44

IFC090, IFC090 i with FF

Troubleshooting

IFC090 FF 01/2006

KROHNE Messtechnik GmbH & Co. KG · Ludwig-Krohne-Str. 5 ‚ D-47058 Duisburg

Tel.: 0203-301 309 ‚ Fax: 0203-301389 · E-mail: [email protected]

Other possibly occuring statuses correspond to the PRIMARY_VALUE statuses. See 6.1.3 for details.

6.1.4 Integrator

Blocks

The follwowing table lists the block errors documented by the Integrator Blocks:

Error Code

Description

Action

Out of Service

Target mode is set
to ‘Out of Service
(OOS)’

If Automatic is the desired target mode, set MODE_BLK target
to ‘Automatic (Auto)’

Memory Failure

Device

memory

failure.

Replace signal converter or contact KROHNE service.

6.2

Transducer Block XD_ERROR

The Transducer Block has a special error parameter XD_ERROR. If this error displays the error ‘Electronics

failure’, the AD converter of the sensor seems to have problems. Please contact the KROHNE support team, if

this error remains.

6.3

Measurement Value Status

Besides the problems that are displayed in the BLOCK_ERR parameter of each block, the status of the

Transducer Block parameter PRIMARY_VALUE (the raw measurement value obtained from the device) can

provide further information on possible reasons for measurement problems.

The following table lists possible indications:

PRIMARY_VALUE
state

Description

Action

BAD:OutOfService:Not
Limited

Target mode is set
to ‘Out of Service
(OOS)’

If Automatic is the desired target mode, set MODE_BLK target
to ‘Automatic (Auto)’

BAD:DeviceFailure:Not
Limited

AD converter fatal
error.

Contact the KROHNE support team.

BAD:SensorFailure:Low
Limited

Measured value is
below the sensor
range.

Check instrument parameters and correct it if necessary. Error
messages deleted automatically.

BAD:SensorFailure:Hig
hLimited

Measured value is
above the sensor
range.

Check instrument parameters and correct it if necessary. Error
messages deleted automatically.

Pipe is empty.

Fill the pipe with media.

Uncertain:NonSpecific:
NotLimited

AD converter
warning.

Check instrument parameters and correct it if necessary. Error
messages deleted automatically.

Uncertain:NonSpecific:
Constant

Current mode is
‘Manual (Man)’

If Automatic is the desired mode, set MODE_BLK traget to
Automatic (Auto)’.

Uncertain:SensorConve
rsionNotAccurate:NotLi
mited

AD converter
overshooted.

Check instrument parameters and correct it if necessary. Error
messages deleted automatically.