KROHNE IFC 010 Converter HART User Manual

Page 15

Advertising
background image

HARTÆ Communications Protocol

IFC010 Command Specification Rev: 1.0

Release Date: 29 May 1997

© 1997 KROHNE Mefltechnik D-47058 Duisburg

Page

15

COMMAND-SPECIFIC RESPONSE CODES

0

No Command-Specific Errors

1-2

Undefined

3

Passed Parameter too Large

4

Passed Parameter too Small

5

Too Few Data Bytes Received

6

Undefined

7

In Write Protect Mode

8

Update Failure

9-113

Undefined

114

Range Exceeded Max and Was Corrected

115

Range Surpassed Min and Was Corrected

116

Both Range and Pulse Value Corrected

117-127

Undefined

8.5

Command #134

Read Low Flow Cutoff Control And Values

Reads low flow cutoff parameters (Menu 1.3).

REQUEST DATA BYTES

NONE

RESPONSE DATA BYTES

#0

#1

#2

CUTOFF

CUTOFF

CUTOFF

CONTROL

ON

OFF

VALUE

VALUE

Data Byte #0

Cutoff control variable:
0 - cutoff option is switched Off
1 - cutoff option is switched On

Data Byte #1

Cutoff On Value, in 1/10% of the Primary Variable Upper Range Value (say, 5% will result
in a value of 50 on a bus). When cutoff is switched Off (Data Byte #0 = 0), the instrument
still uses the cutoff option, and the defaulted On Value is returned (that is not stored in
EEPROM!). The latter might be either 0.1% or 1% depending on the PulseType value (refer
to the instrument Operating Instructions.

Data Byte #2

Cutoff Off Value, in 1/10% of the Primary Variable Upper Range Value. When cutoff is
switched Off (Data Byte #0 = 0), the instrument still uses the cutoff option, and the defaulted
Off Value is returned (that is not stored in EEPROM!). The latter might be either 0.2% or 2%
depending on the PulseType value (refer to the instrument Operating Instructions.

COMMAND-SPECIFIC RESPONSE CODES

0

No Command-Specific Errors

1-127

Undefined

8.6

Command #135

Write Low Flow Cutoff Control And Values

Writes low flow cutoff parameters.

REQUEST DATA BYTES

#0

#1

#2

CUTOFF

CUTOFF

CUTOFF

CONTROL

ON

OFF

VALUE

VALUE

Advertising