Operation, 6 status messages and diagnostic information, 2 deleting error messages in the menu "quick setup – KROHNE IFC 050 Converter EN User Manual

Page 64

Advertising
background image

6

OPERATION

64

IFC 050

www.krohne.com

08/2013 - 4002184002 - MA IFC 050 R02 en

6.5.2 Deleting error messages in the menu "quick setup"

6.6 Status messages and diagnostic information

Operational faults in device

INFORMATION!
For the detailed list of the possible error messages refer to Status messages and diagnostic
information on page 64

.

Key

Display

Description and setting

>

quick setup

Press and hold for 2.5 s, then release the key.

>

language

-

3 x ↓

reset

-

>

reset errors

-

>

reset?

no

-

↓ or ↑

reset?

yes

-

^

reset errors

Error has been reset.

3 x ^

Measuring mode

-

Messages on the display Description

Actions

Status: F _ _ _ _ _

Operational fault in device, mA output ≤
3.6 mA or set fault current (depending
on the seriousness of the fault), status
output open, pulse / frequency output:
no pulses

Repair necessary.

F error in device

Error or failure of device. Parameter or

hardware error. No measurement possible.

Group message, when one of the following

or some other severe error occurs.

F IO 1

Error, operational fault in IO 1. Parameter

or hardware error. No measurement

possible.

Load settings (Fct. C4.6.3) (backup 1,

backup 2 or factory settings). If status

message still does not disappear, replace

electronic unit.

F parameter

Error, operational fault of data manager,

electronic unit, parameter or hardware

error. Parameters no longer usable.

F configuration

Invalid configuration: display software, bus

parameter or main software do not match

existing configuration.

If device configuration unchanged:

defective, replace electronic unit.

F display

Error, operational fault in display.

Parameter or hardware error. No

measurement possible.

Defective, replace electronic unit.

F sensor electronic

Error, operational fault in sensor

electronics. Parameter or hardware error.

No measurement possible.

Defective, replace electronic unit.

F sensor global

Data error in the global data of the

measuring sensor electronic equipment.

Load settings (Fct. C5.6.3) (backup 1,

backup 2 or factory settings). If status

message still does not disappear, replace

electronic unit.

F sensor local

Data error in the local data of the measuring

sensor electronic equipment.

Defective, replace electronic unit.

F field current local

Data error in the local data of the field

current supply.

Defective, replace electronic unit.