Operation – KROHNE H250 M8-M9-M10 EN User Manual

Page 51

Advertising
background image

OPERATION

6

51

H250

www.krohne.com

04/2011 - 4000269302 MA H250-R02 en

6.5.3 Menu explanations

Level Designation

Selection /
Input

Explanation

1.1.1

OUTPUT B1

INACTIVE

FLOW.VAL B1

Flow value switching point.

The switching point is entered in flow units.

If the current flow value exceeds this set switching point, then output B1

is activated.

Note:

The function NC or NO can be selected using menu 3.3.

COUNTER.VAL

B1

Counter value switching point.

Each positive number can be set here.

If the counter exceeds this value, then output B1 is activated.

Note:

The function NC or NO can be selected using menu 3.3.

1.1.2

OUTPUT B1

HYST.B1

Hysteresis setting for the flow value switching point.

Value range 0...switching point.

Example, if a switching value of 200 is set under 1.1.1, then a hysteresis

value of 0...200 can be set here.

If the value 0 is entered here, then this output does not have hysteresis.

If the value 20 is entered here, then the output functions as follows:

If the current flow value exceeds the value 200, then the output switches

3.

If the current flow value drops below the hysteresis value of 180, then the

switching output returns to its normal state 4.

Note:

To invert the operating method, use menu 3.3 to set the output from NO

1 to NC 2 or vice versa.

This function is not activated on the counter switching point.

1.2.1

OUTPUT B2

INACTIVE

FLOW.VAL B2

see FLOW.VAL B1

COUNTER.VAL

B2

see COUNTER.VAL B1

PUL. VAL B2

B2 = pulse output

Note:

Settings under menu 3.6 pulse duration and

3.7 Pulses/unit

1.2.2

OUTPUT B2

HYST.B2

See HYST. B1

1.3

DISPLAY

FLOW RATE

COUNTER

FLOW&COUNT

PERCENT

MA_H250_R02_en_269302_PRT.book Page 51 Tuesday, April 12, 2011 10:17 AM

Advertising