Contact function, Electrical connection, Setting – KROHNE H250 Reed MS14 EN User Manual

Page 2

Advertising
background image

11/2012

MA H250-MS14-Ex-II2G-AD-R01-en 4000602001

Electrical connection

For the electrical connection, first unscrew the cover. Because the reed contact is potential-free, the

terminals are not labeled. The intrinsically safe equipment may only be connected to separated intrinsically

safe circuits.

Setting

The MS14 is fixed to the flowmeter with a special support and is adjustable over the full measuring

range.

The limit switch MS14 is a potential-free, bistable reed contact. It can be operated optionally as an

NO or NC contact. Unless specified otherwise, it is supplied by the factory as an NO contact.



Untighten support screw (1).



Set operating point.



Tighten support screw (1).

Contact function

Two contact functions are available for limit monitoring:

Closing, at decreasing flow:

Arrow on reed cartridge points away from measuring cone.

Closing, at increasing flow:

Arrow on reed cartridge points towards measuring cone.

The contact function can be changed by changing over the reed cartridge screwed into the

cartridge case: