KROHNE ALTOSONIC III EN User Manual

Page 21

Advertising
background image

Installation and operating instructions

21 ALTOSONIC

III



Error message

Description of error message

OIML error
(set function
3.10.1 to CUST.

ERR)

What to do?

CURR > MAX

Current output overflow (> 22 mA)

No

Check flow velocity

DSP

Digital signal processor (DSP) internal
error

Yes

Only checked at power-up. Switch
off and on the flowmeter. If the
error still exists, contact KROHNE
representative

EE MENU

Menu parameters corrupted

Yes

Contact KROHNE representative

EE SERVICE

Service parameters internal error

Yes

Contact KROHNE representative

EMPTY PIPE

Measuring tube not completely filled,
flow reading to 0, error on all 3 paths.

Yes

Fill measuring tube completely

FLOW > MAX

Measuring range overflow (flow > 2 x
Q

max

)

No

Check flow velocity

FRONT END

Front end internal error

Yes

Only checked at power-up. Switch
off and on the flowmeter. If the
error still exists, contact KROHNE
representative

RESTART

Flowmeter restarted

No

Reset errors

UNRELIABLE

Flow data disturbed, same as right
compass field (4)

Yes

Check flow conditions

PATH 1

Measuring path 1 error

Yes

Check flow conditions

PATH 2

Measuring path 2 error

Yes

Check flow conditions

PATH 3

Measuring path 3 error

Yes

Check flow conditions

PULS > MAX

Pulse output overflow (> 120 %)

No

Check flow velocity

UP2

µP2 internal error

Yes

Contact KROHNE service

VOS path 1

Deviation > 5% from average VOS of
3 measuring paths

Yes

Check VOS of each path

VOS path 2

Deviation > 5% from average VOS of
3 measuring paths

Yes

Check VOS of each path

VOS path 3

Deviation > 5% from average VOS of
3 measuring paths

Yes

Check VOS of each path

Flow calc.

Algorithm check error

Yes

Reset error

Cal data

Yes

Contact KROHNE representative

System stp

System error

Yes

Power on/off. Contact KROHNE
representative when message
remains

Function 0.00.03 Reset Totalizer
Reset display totalizer(s). Only available when Function 3.07.08 RST ENABLE is YES. Note that all totalizer values are
reset.

3.11.2 Main menu 1.00.00

Operation

The functions in this menu are a subset of Main menu 3.00.00 Installation, and are selected in this menu as most
commonly used functions for a quick installation. Note that parameters set in these functions are automatically set in
both menus.

3.11.3 Main menu 2.00.00

Test functions

This menu is for testing the display, the in- and outputs and for information on hard- and software numbers. See
chapter on functional checks and service.

Advertising