KROHNE ALTOSONIC V EN User Manual

Page 31

Advertising
background image

ALTOSONIC V

Operating Manual rev02 E

7.30850.34.00

Page 31 of 106

4.8.1

Controls menu: F2 API settings window

In this window the configuration can be made for calculating the standard volume /mass.

The green arrows

represent the current settings per option. The red arrow

is the option cursor.

CALCULATION option is configurable:

1. Disable, no standard volume or mass is calculated
2. Standard volume/mass by API standards
3. Mass measurement by input of proces density.

TEMPERATURE STANDARD:

When the CALCULATION option is 2, the used temperature standard is selectable between 0-30°C or

equivalent in °F. If the temperature standard is changed, the input limits for the density standard per

fluid type also change to default and have to be configured as desired.

DENSITY STANDARD BY:

When the CALCULATION option is 2 then the method to establish the density standard is configurable:

1. Fill in manually value for the density standard manually in this window. Additional only proces

temperature and pressure must be measured.

2. Calculated from Densitometer density. The density standard is calculated by iteration of the

measured density (on frequency or AD input). Additional proces and densitometer temperatures

and pressures must be measured.

3. On AD input. Density standard on an AD input. Additional only proces temperature and pressure

must be measured and the temperature standard must be set according to input density standard.

FLUID TYPE:

When the CALCULATION option is 2 then the used fluid type is configurable. Each fluid type has its

own density standard limits.

Advertising