START-UP

5

51

OPTIBAR PC 5060 C

www.krohne.com

04/2014 - 4003437201 - MA OPTIBAR PC 5060 C R01 en

Out of specification

Maintenance required

Code
Text message

Cause

Action or description

S600

Inadmissible electronics

temperature

Temperature of the electronics in

the non-specified section

Check ambient temperature

Isolate electronics

Use device with higher

temperature range

S603

Inadmissible operating voltage

Operating voltage below specified

range

Check electrical connection - if

necessary, increase operating

voltage

Code
Text message

Cause

Action or description

M500

Error with the reset delivery

status

Saved reset delivery status is

incorrect

Send device for repair

M501

Error in the non-active

linearization table

Index markers are not

continuously rising, for example

illogical value pairs

Check linearisation table,

Delete table/Create new

M502

Error in the event memory

Hardware error in EEPROM

Exchange the electronics

Send device for repair

M504

Error on a device interface

Hardware defect

Check connections

Exchange the electronics

Send device for repair

M507

Error in the device settings

Error during setup

Error when carrying out a reset

Repeat setup,

Repeat reset