7.4 Faults and symptoms during start-up and process flow measurement

Before contacting Krohne Service, please read through the

following tables. THANK YOU !

Most faults and symptoms occurring with the compact flowmeters can be eliminated
with the aid of the following tables.

For greater clarity, faults and symptoms in the tables are divided into 2 parts and
various groups.

Part 1

Signal converter IFC 010 B (B = basic version), without display
and without HHT or CONFIG user program (see Sect. 6.1)
Groups:

LED

LED display (status messages)

I

Current output

P

Pulse output

LED / I / P LED display, current output and pulse output

Part 2

Signal converter IFC 010 D (D = display version) and
Signal converter IFC 010 B (B = basic version), without display
but with CONFIG user program (see Sect. 6.1)
Groups:

D

Display

I

Current output

P

Pulse output

S

Status indication output

D / I / P / S LED display, current output,

status output

Part 1

Converter IFC 010 B (B = basic version), without display
and without HHT or CONFIG operator program

Group LED

Fault / Symptom

Cause

Remedial action

LED 1

LED flashes red/green

Overranging of A/D converter,

Reduce flowrate; if unsuccessful,

current or pulse output

test as described in Sect. 7.5.

Measuring tube drained,

Fill measuring tube.

A/D converter overranged.

LED 2

LED flashes red

Fatal Error, hardware and/or

Replace signal converter

software error

(see Sect. 8.4) or contact
Krohne Service.

LED 3

cyclic flashing of red LED,

Hardware fault, Watchdog

Replace signal converter

approx 1 second

activated.

(see Sect. 8.4) or contact
Krohne Service.

LED 4

LED shows red continuously

Hardware fault

Replace signal converter
(see Sect. 8.4) or contact
Krohne Service.