KROHNE CORIMASS G-MFM 4085 EN User Manual

Page 48

Advertising
background image

48

To set the Time Constant :


Begin from measuring mode.


This filtering only applies to mass and volume flow readings and any outputs that use them. The
mass totaliser is independent of the time constant.


The standard range for the time constant is 0.5 to 20 seconds. For faster response times, a
range of 0.2 to 20 seconds is available as a factory set option.

Time constant characteristic


5.4

Programming the display for measurement values (Fct. 1.2. and 3.2.)


The following measurement functions can be displayed :


Key

Display

line 1

line 2


→→→

Fct. 1.1.(1)

ZERO SET

↑↑

Fct. 1.1.(3)

TIME. CONST.

(0)4.0

TIME C. S.

Edit the time constant in

seconds in range 0.5 to 20.

Fct. 1.1.(3)

TIME.CONST

4x

Fct. 1.2.1

CYCL.DISP

Fct. 1.2.2

STATUS MSG.

Fct. 1.2.3

MASS FLOW

Fct. 1.2.4

MASS TOTAL

Fct. 1.2.5

DENSITY

Fct. 1.2.6

TEMPERAT.

Fct. 1.2.7

VOLUME.FLOW

Fct. 1.2.8

VOL.TOTAL

Advertising