Conical silo nozzles, False readings and traction on the cable probes


Caution

High traction forces :

1

We recommend that the probe should not be
anchored to avoid excessive traction loads on
the cable.

Bending and traction:

2

Position the connection on the roof at ½
radius of the tank and with minimum nozzle
height. This will avoid damage due to bending
and traction during emptying.



Traction forces during emptying cycles for powder applications

Traction load is dependent upon the height and shape of the tank, product particle size & density,
and the rate at which the tank is emptied. The table below gives the load at which cable probes will
support.


Cable maximum design load, traction
Probe Maximum

Load

Single cable Ø8 mm or Ø0.3”

3.5 T or 7700 lb



Traction on cable according to product (approximate value in metric tons)

Probe Length / m (ft)

Probe used

Material

10 (33) 20

(65.5) 24

(79)

Cement

1.0 T or 2200 lb

2.0 T or 4410 lb

2.4 T or 5290 lb

Single cable Ø8mm or
Ø0.3”

Flyash

0.5 T or 1100 lb

1.0 T or 2200 lb

1.2 T or 2650 lb