(5) Connect the test leads to the circuit under test so the instrument is in the series.

(6) Power on the circuit under test.

(7) Measured value is indicated on the display.

6-2-4 AC Current Measurement (up to 10A)

CAUTION

● For safety sake, the measuring time on 10A range should be 15 seconds or less. Interval time between 2 measurements should be greater than 15 minutes.

If you carry on making measurement continuously over 15 seconds, or make measurement in short interval time, it may cause error of measurement and

instrument damage.

(1) Insert the black test lead plug into the COM terminal and the red test lead plug into the 10A terminal.

(2) Set the Function Selector Switch to "A" position.(Then, "DC" , "Auto", and "A" symbols are indicated on the display.)

(3) Set the instrument to AC mode by pressing "SELECT" Key.(Then, "AC" symbol is indicated on the display.)

(4) Power off the circuit under test.

(5) Connect the test leads to the circuit under test so the instrument is in the series.

(6) Power on the circuit under test.

(7) Measured value is indicated on the display.

6-3 Resistance Measurement (Ω/ Diode check/ Continuity Check)

DANGER

● To avoid the danger of getting electrical shock, never make measurement of the circuit in which electric potential exists.

● Do not make measurement when opening the battery cover and the instrument case.

6-3-1 Resistance Measurement

(1) Insert the black test lead plug into the COM terminal and the red test lead plug into the VΩHz terminal.

(2) Set the Function Selector Switch to "Ω" position.(Then, "Auto" and "MΩ" symbols are indicated on the display.)

Make sure that the "O.L" symbol is indicated on the display at this bout, then short the test lead tips and check "0" is indicated on the display.

(3) Connect the test leads to both ends of the resistance under test.

Measured value is indicated on the display.

Note) Even if short the test lead tips, indicated value may not be "0". But this is because of the resistance of test leads and not a failure.

In that case, by pressing "REL" key, "0" will be indicated.

6-3-2 Diode Check

(1) Insert the black test lead plug into the COM terminal the red test lead plug into the VΩHz terminal.

(2) Set the Function Selector Switch to "Ω" position.(Then, "Auto" and "MΩ" symbols are indicated on the display.)

(3) Press the "SELECT" Key twice and set the instrument to Diode Check mode.

(Then, "

" and "V" symbols are indicated on the display.)

Make sure that the ".OL" symbol is indicated on the display at this bout, then short the test lead tips and check "0" is indicated on the display.

(4) Connect the black test lead to the cathode side of the Diode and the red test lead to the anode side of the Diode.

Forward voltage of Diode is indicated on the display.

(5) Connect the black test lead to the anode side of the Diode and the red test lead to the cathode side of the Diode.

Normally, ".OL" symbol is indicated on the display.

Conclusion : Diode is OK if the instrument complies with above items(4) and (5).

Note) Open-circuit voltage between measuring terminals is approx.2.8V.(measuring current approx. 0.4mA)

6-3-3 Continuity Check

(1) Insert the red test lead plug into the VΩHz terminal and the black test lead plug into COM terminal.

(2) Set the Function Selector Switch to "Ω" position.(Then, "Auto" and "MΩ" symbols are indicated on the display.)

(3) Press the "SELECT" Key one time and set the instrument to Continuity Check mode.