KYORITSU 1030 User Manual

Kew 1030, Instruction manual, Pen type digital multimeter

Advertising
background image

4−2 Protective cover

CAUTION

¡Do not apply excessive force to the Test pin and the Protective cover.
¡Be careful not get hurt by the tip of the Test pin when setting or

releasing the Protective cover.

Use the Protective cover to cover the Test pin when carrying or
storing the instrument.

1

Method of setting the Protective cover

Pinch the tip of the Protective cover, and pull it towards the tip direction.
Then turn it 90 degrees as shown in the figure below to match the
marks on the cover and on the instrument body.

Continuity
Check

50nF

500nF

5uF

50uF

100uF

5Hz

50Hz

500Hz

5kHz

50kHz

200kHz

○Read through the following safety instructions contained in this

manual before using the instrument.

DANGER

¡Never make measurement on a circuit in which electrical potential to

ground over 600V exists.

¡Do not attempt to make measurement in the presence of flammable

gasses. Otherwise, the use of the instrument may cause sparking,

which can lead to an explosion.

¡Never attempt to use the instrument if its surface or your hand is wet.

Otherwise, you may get electrical shock.

¡Never open the Bottom case and Battery cover during a measurement.

WARNING

¡

Never attempt to make any measurement if any abnormal conditions,

such as broken case and exposed metal parts are present on the

instrument or test lead.

¡

Do not install substitute parts or make any modification to the

instrument. Return the instrument to your local Kyoritsu distributor for

repair or re-calibration.

CAUTION

¡

Always set the Function switch to the appropriate position before

making measurement.

¡

Do not expose the instrument to the direct sun, high temperatures and

humidity or dew.

¡

This instrument is designed for in-door use. It can be used under the

temperature between 0℃ and 40℃ without impairing its safety

characteristics.

¡

This instrument doesn't have dust/water-proof construction. Do not use

the instrument in dusty area or where it easily gets wet. It may lead to

failure of the instrument.

¡

Set the Function switch to "OFF" position after use. Remove the

batteries if the instrument is to be stored and will not be in use for a

long period.

○Measurement categories (Over-voltage categories)

To ensure safe operation of measuring instruments, IEC61010

establishes safety standards for various electrical environments,

categorized as CAT.l to CAT.IV, and called measurement categories.

These are defined as indicated below.

Higher-numbered categories correspond to electrical environments with

greater momentary energy, so a measuring instrument designed for

CAT.III environments can endure greater momentary energy than one

designed for CAT.II.

INSTRUCTION MANUAL

Thank you for purchasing our instrument KEW1030.
Before using the instrument,read this manual thoroughly to obtain the
maximum performance of this instrument and ensure the correct
measurement.

3−1 Accuracy [ guaranteed at temperature & humidity: 23±5℃,45∼85% RH(*1) ]

KYORITSU ELECTRICAL INSTRUMENTS

WORKS, LTD.

Function

Range

Accuracy

PEN TYPE DIGITAL MULTIMETER

KEW 1030

This instrument is a pen-type digital multimeter and can measure:
AC/DC voltage, resistance, capacitance and frequency/duty ratio. It also
provides continuity check and diode check functions.

¡

Designed to meet the following safety standards.
IEC61010-1 measurement category (CAT.) III 600V
IEC61010-031 (for hand-held Probe assemblies)

¡

Double molded main body and Function switch provide comfortable
single handed grip.

¡

Penlight illuminates brightly the point to be measured.

¡

Backlight LCD is highly visible, even in darkness.

¡

REL function to check the difference (DC.V/ CAP).

¡

Auto-power-off function to save battery.

¡

Data hold function

¡

All ranges including Ohm range are protected against overload
voltage of 600V.

¡

Test lead is wrapped in its rear side compartment without difficulty.

¡

Test pin can be covered by a unique cover mechanism for safety.

Max. input voltage

ACV

Auto-ranging(*2)

DCV

Auto-ranging(*2)

400mV

4V

40V

400V
600V

Ω

Auto-ranging

400Ω

4kΩ

40kΩ

400kΩ

4MΩ

40MΩ

Diode check/

Continuity Check

Capacitance

Auto-ranging

Frequency

Auto-ranging

DUTY(pulsewidth/

pulse cycle)

DC 600V

AC 600Vrms(sin)

0.1∼99.9%

±2.5%rdg±5dgt(Accuracy is

guaranteed up to 10kHz.)

±1.3%rdg±5dgt (50/60Hz)
±1.7%rdg±5dgt (∼400Hz)
±1.6%rdg±5dgt (50/60Hz)
±2.0%rdg±5dgt (∼400Hz)

±0.8%rdg±5dgt

±1.0%rdg±5dgt

±1.0%rdg±5dgt

±2.5%rdg±5dgt

Diode check Test voltage:approx. 0.3V∼1.5V

Buzzer sounds when
resistance is 120Ω or less.
±3.5%rdg±10dgt

±4.5%rdg±5dgt

±3.5%rdg±5dgt

±0.1%rdg±5dgt
Measurable input:

1.5Vrms or more

Note:

Following abbreviations are used in above table.

¡

rdg is an abbreviation of "reading", and it means the indicated value at a

measurement.

¡

dgt is an abbreviation of "digit", and it means the figure to be displayed

at the rightmost digit.

¡

(*1): Except for 40MΩ range at Ohm function.

¡

(*2): At Voltage function, the Auto-ranging function is released by pressing

the SELECT key. To measure a voltage again, turn the Function switch to

the "OFF" position once. Then set it to the Voltage function again.

User must refer to the explanations in the instruction manual.

F

Instrument with double or reinforced insulation

AC

---

DC

1.Safety warnings

2.Features

lncoming wire

lnterior wiring

Transtormer

Socket

3.Specification

4V

40V

400V
600V

○This instrument has been designed, manufactured and tested according

to IEC 61010: Safety requirements for Electronic Measuring apparatus,
and delivered in the best condition after passed the inspection.
This instruction manual contains warnings and safety rules which
must be observed by the user to ensure safe operation of the
instrument and retain it in safe condition. Therefore, read through
these operating instructions before using the instrument.

WARNING

¡

Read through and understand the instructions contained in this manual
before using the instrument.

¡

Save and keep the manual at hand to enable quick reference
whenever necessary.

¡

The instrument is to be used only in its intended applications.

¡

Understand and follow all the safety instructions contained in the
manual.

¡

The RESPONSIBLE BODY shall be made aware that, if the equipment is
used in a manner not specified by the manufacturer, the protection
provided by the equipment may be impaired.

○The symbol

indicated on the instrument means that the user must refer

to the related parts in the manual for safe operation of the instrument.
Be sure to carefully read the instructions following each

symbol in the

manual.

DANGER :is reserved for conditions and actions that are likely to

cause serious or fatal injury.

WARNING:is reserved for conditions and actions that can cause

serious or fatal injury.

CAUTION :is reserved for conditions and actions that can cause

injury or instrument damage.

○ Please refer to following explanation of the symbols used on the

instrument and in this manual.

CAT.Ⅰ:Secondary electrical circuits connected to an AC electrical

outlet through a transformer or similar device.

CAT.Ⅱ:Primary electrical circuits of equipment connected to an AC

electrical outlet by a power cord.

CAT.Ⅲ:Primary electrical circuits of the equipment connected directly

to the distribution panel, and feeders from the distribution panel

to outlets.

CAT.Ⅳ:The circuit from the service drop to the service entrance, and to

the power meter and primary over-current protection device

(distribution panel).

3−2

General specification

¡

Method of operation : △∑ method

¡

Display

: Liquid crystal display (max. 3999 counts)/Units/

Marks

¡

Over-range indication: "

OL

" displayed when exceeding the measuring

range. (except for AC/DC 600V range)

¡

Range switching

: Fully-automatic range

(Single range is available at Continuity, Diode check

and DUTY range.)

Range shifts to upper range:4000 counts or more.

Range shifts to lower range:less than 360 counts.

¡

Sample rate

: twice per second

¡

Functional construction : OFF/ ACV/ DCV/Ω/ Capacitance

¡

Key

: HOLD/ Hz/ DUTY/

/

/

RELΔ (only at DCV and Capacitance ranges)

¡

Power source

: Button type battery LR44(SR44)1.5V x 2

¡

Low battery warning : "

" mark is displayed at 2.4V±0.2V or less.

¡

Dimension

: 190(L) x 39(W) x 31(D)mm

¡

Weight

: Approx. 100g (including batteries)

¡

Location for use

: Altitude up to 2000m, in-door use

¡

Operating temperature : 0∼40℃, relative humidity 85% or less

& humidity range

(no condensation)

¡

Storage temperature : -20∼60℃, relative humidity 85% or less

& humidity range

(no condensation)

¡

Accessories

: Carrying case x 1

Button type battery LR44(1.5V) x 2

Instruction manual x 1

¡

Standards (Safety)

: ・IEC/EN 61010-1:2001

Measurement category (CAT.) III 600V

Pollution degree 2

・IEC/EN 61010-031:2002

(EMC)

: ・EN 61326:2001

3−3

Electrical characteristics

¡

Temperature &

: 23℃±5℃, relative humidity 85% or less

humidity range

(no condensation)

(guaranteed accuracy)

¡

Supply voltage range : 3.4V till the "

" mark is displayed.

(guaranteed accuracy)

¡

Insulation resistance : 10MΩ or more/ DC1000V

(between electrical circuit and case enclosure)

¡

Withstand voltage

: AC5.55kVrms, sine wave (50/60Hz for 1 min.)

(between electrical circuit and case enclosure)

¡

Overload protection : 720V (AC/DC) for 10 sec. at voltage function

(Over-voltage protection) 600V (AC/DC) for 10 sec. at all functions

other than voltage function

¡

Rated supply voltage : DC3.0V

¡

Rated power

: Approx. 4mVA (when battery voltage is 3.0V)

¡

Max. rated power

: Approx. 30mVA (when lights are on)

¡

Continuous operating: Approx. 80 hours (DCV measurement)

time

Approx. 15 hours (A operation; turning the light on

for 10 sec. and off for 20 sec., is repeated.)

4.Instrument layout

4−1 Instrument layout

①Test pin (input terminal (+); red)
②Test lead (input terminal (−); black)

:Connected to the negative (−) side or the earth of the circuit.

③ Protective cover

:Covering the Test pin for safety purpose.

④ Penlight
⑤ Barrier
⑥ Function switch

■OFF :Power off (Battery will not be wasted.)

AC voltage (ACV)

Frequency (Hz)

DUTY(%)

Switches by pressing the "SELECT" key.

DC voltage (DCV)

RELΔ (relative value display)

Switches by pressing the "SELECT" key.

Ω Resistance

Diode check

Continuity check

Switches by pressing the "SELECT" key.

Capacitance

RELΔ (relative value)

Switches by pressing the "SELECT" key.

V

V

■ LIGHT:Turning on the Penlight. Set the Function switch to this

position first, and then turn it to any desirable function position.
Then the Penlight turned on and illuminates the test point.
(Measurement cannot be performed in this switch position.)

⑦ HOLD key

・Freezing the indicated value.
・Turning on the LCD backlight. (Press this key at least 2 sec.)

⑧ SELECT key

・Switching the measurement modes.(

/ Hz/ DUTY andΩ/

/

・Enable/ Disable the RELΔ function. (Only at DCV/ Capacitance)

⑨ LCD indication

V

Low battery warning

Auto-ranging

DC

AC

Polarity

Hold

Diode check

Continuity check

RELΔ

Unit

Set

Release

2

Method of releasing the Protective cover

Pinch the tip of the Protective cover, and pull it towards the tip direction.
Then turn it 90 degrees as shown in the above figure. Then the cover is
stored automatically and the Test pin (positive terminal) appears.

3

Press the SELECT key to display a REL value (relative value).

Press this key and store the initial measured value. After that, the
difference between the stored value and the measured value is
displayed on the LCD. Auto-ranging function doesn't activate when this
function is enabled. The first selected range will be held. The relative
measurement is allowed in the following range.
* Measuring range = Full scale value at a range − initial value

Press this key again or switch the measurement function to others to
release the REL function.

¡Following measurements can be done

by pressing the SELECT key.

DCV RELΔ (relative value)

"△" mark is displayed

on the LCD.

WARNING

¡To avoid getting electrical shock, be sure to remove the measuring

terminals from the equipment under test; set the Function switch to

OFF position before replacing batteries.

DANGER

¡Do not mix new and old batteries. Never mix the different kinds of

batteries.

¡Make sure to install batteries in correct polarity as marked inside.

¡Be sure to fasten the Battery case-fixing screws after the battery

replacement.

CAUTION

¡Dispose the used batteries according to the rules, which are defined by

each community.

①Set the Function switch to OFF position.

②Loosen one Battery cover-fixing screw, and remove the Battery cover.

③Replace the batteries with new ones. Make sure to install batteries in

correct polarity as marked inside. Always replace all two batteries with

new ones at the same time.

④Put the Battery case at the original position, and fasten the screws.

Note
¡When the forward voltage of diode is out of the range of 0.3V∼1.5V,

measurement may not be done. (Zener diode, LED and etc.)

3

Press the SELECT key and select the Frequency range to measure a
frequency. In this case, the unit "Hz" is displayed on the LCD.

¡Following measurements can be done

by pressing the SELECT key.

ACV

Frequency

DUTY ratio

4

Press the SELECT key and select the DUTY ratio range to measure
a DUTY ratio (Pulse width/ Pulse cycle). In this case, the unit "%" is
displayed on the LCD.

Note

¡At ACV function, a few dgts may remain displayed on the LCD after

removing the input.

¡Connect the test lead (minus terminal) to the earth side of the circuit

under test. When the circuit under test does not have the earth, any
connection is allowed.

¡At Frequency and DUTY ratio measurement, the measurable min.

input is approx. 1.5Vrms.

¡Auto-ranging (AUTO)

A function to automatically select the appropriate measurement range

based on the input signal. The "

AUTO

" mark is displayed on the LCD

while this function is activated. This function is not available in Diode

check, Continuity check and Duty ratio measurements. The "

AUTO

"

mark is not displayed.

¡Hold function ( )

A function to freeze the measured value on the LCD. (Not available in

Frequency measurement)

The " " mark is displayed on the LCD when the HOLD key is pressed.

Then the measured value is frozen. Press this key again or switch the

measurement function to others to release the Hold function.

¡REL function ( Δ )

A function to display the difference between the measured values

(relative value) on the LCD at DCV and Capacitance functions. The "Δ"

mark is displayed on the LCD when the HOLD key is pressed. Then the

value being measured is stored. After that, the difference between the

stored value and the measured value is displayed on the LCD. Press this

key again or switch the measurement function to others to release the

REL function.

¡Auto-power-off function

A function to turn off the instrument when 30 min. have elapsed after the

Function switch is switched from OFF to the other measurement function.

Press the HOLD key again or switch the measurement function to others

to restore from the Auto-power-off state.

¡Over-range indication

When the measured value exceeds the max. indication range, "

OL

" is

displayed on the LCD. (This indication is not displayed at AC/DC 600V

range.) This indication is not displayed while the Hold function is

activated.

5.Functions

1

Set the Function switch to"

"position.

2

Connect the Test pin and test lead to AC circuit as shown in the figure
below to measure AC voltage (ACV).

V

6−1 AC voltage(ACV), Frequency and DUTY ratio measurement

DISTRIBUTOR

Kyoritsu reserves the rights to change specifications or designs
desscribed in this manual without notice and without obligationss.

KYORITSU ELECTRICAL
INSTRUMENTS
WORKS, LTD.

No.5-20,Nakane 2-chome,Meguro-ku,
Tokyo, 152-0031 Japan
Phone:81-3-3723-0131 Fax:81-3-3723-0152
URL:http://www.kew-ltd.co.jp
E-mail:[email protected]
Factories:Uwajima&Ehime

04-11

92-1634A

4−3 Method of storing the test lead

Test lead is stored in the rear side compartment of the instrument.
Cord is winded around the cord holder.

¡Low battery warning ( )

When the battery voltage drops to 2.4V±0.2V or less, the " " mark is
displayed on the LCD.

¡Penlight

Set the Function switch to "LIGHT" position
to turn on the Penlight. Turn the switch
to any desirable function position .
(Measurement cannot be performed when
the switch is in "LIGHT" position.) Turn the
switch to "OFF" position to turn off the light.

¡LCD backlight

The LCD backlight lights up by pressing down the HOLD key at any
measurement function other than OFF at least 2 sec..
Press down this key again at least 2 sec. or turn the Function switch to
OFF once to turn off the light.

Note
¡Penlight and LCD backlight are not turned off automatically. Be sure to

turn them off when they are not in use.

¡When turning on/ off the LCD backlight, the "

" mark is displayed on

the LCD and the Hold function is activated.
Press the HOLD button for a while to release the function and perform
the next measurement.

DANGER

To prevent electrical shock to person and damage to the instrument,
following instructions must be observed.
¡The max. rated voltage to ground is AC/DC600V. Never attempt to

make measurement on a circuit in which electrical potential to the
ground exceeding this voltage exists.

¡The max. input voltage is DC600V/AC600Vrms (sin). Never attempt to

make any measurement on a circuit in which electrical potential
exceeding this voltage exists.

¡Do not operate the Function switch during a measurement.
¡Never make a measurement with the Bottom case is removed.
¡Keep your fingers and hand behind the barrier (see 4-1) of the

instrument and test lead.

¡Be careful not to short-circuit the line under test with the metal part of

the instrument or the test lead during a measurement.

¡Never make measurement on an energized circuit at Resistance,

Diode check, Continuity check and Capacitance function of this
instrument.

6.Measurement

6−2 DC voltage(DCV) measurement

1

Set the Function switch to"

"position.

2

Connect the Test pin to the positive (+) side of the equipment under test
and the test lead to the negative (-) side as shown in the figure below.
When test lead is connected to the positive (+) side, the "-" mark is
displayed on the LCD.

V

6−3

Resistance (Ω) measurement, Diode/ Continuity check

1

Set the Function switch to" Ω "position.

2

Connect the Test pin and test lead to the equipment under test as
shown in the figure below.

3

Press the SELECT key to conduct the Diode check.

Connect the Test pin and the test lead to the equipment under
test. When following indication is confirmed, the diode is good.

¡Following measurements can be done

by pressing the SELECT key.

Ω   Diode check

Continuity check

Cathode

Anode

1.Forward direction

(Example)
Forward voltage
is displayed.

Anode

Cathode

2.Backward direction

(Example)
"OL"
is displayed.

4

Press the SELECT key to conduct the Continuity check. Connect the

Test pin and the test lead to the equipment under test.
Buzzer sounds when continuity is ok. (120Ω or less)
Resistance value of 400Ω or less is displayed on the LCD.

Note
¡Indicated value may not be "0" after shorting the tip of the test lead.

However, this is because of the resistance of the test lead and not a
failure.

2

Press the SELECT key to make the indicated value to "0" before

connecting the test lead to the equipment under test.

¡Press the SELECT key to make the

indication to 0.

3

Connect the Test pin and the test lead to the equipment under test as

shown in the figure below.

Note
¡Measuring time varies depending on the capacitance to be

measured.

Capacitance

to be measured

Measuring

time

<4μF

2 sec.

<40μF

7 sec.

<100μF

15 sec.

1

Set the Function switch to "

" position.

6−4 Capacitance measurement(nF , μF)

7.Battery replacement

Battery cover-fixing screw

Button type battery
 LR44(SR44) 1.5V×2pcs

8.Maintenance

¡Cleaning

Use a cloth dipped in water or neutral detergent for cleaning the

instrument.

Do not use abrasives or solvents. Otherwise, instrument get

damaged,deformed or discolored.

①Penlight ON

②Select any

 function to make

 measurement

Measurement function

③Penlight
 OFF