8. BATTERY & FUSE REPLACEMENT

DANGER

Never open the battery compartment cover while making measurement.

To avoid possible electrical shock, disconnect the test probe before

opening the cover for battery and fuse replacement.

Replacement fuse must be have the following rating.

Fast acting type, F 500mA/600V,

φ

6.35

×

32mm

8-1 Battery Replacement

Disconnect test probes from the instrument.

Open the battery compartment cover by unscrewing the metal captive

screw to reveal battery compartment. Always replace all eight

batteries with new ones at the same time.

Battery type: 8 x R6P, 1.5V AA or equivalent

8-2 Fuse Replacement

Disconnect the test probe from the instrument.

Open the battery compartment cover by unscrewing the metal captive

screw to reveal battery compartment and replace the fuse.

Fuse type: 600V/500mA (F) quick acting ceramic fuse 6.35 x 32mm

19

7. FUNCTIONS

7-1 TRAC-LOK MODE (Model 3007A)

TRAC mode : Measurement can be conducted while the test

button is

being pressed.

When making continuous measurement, select this mode.

LOK mode : When the test button is pressed, measurement can be conducted

only once, and output is stopped, then automatically discharged.

This allows to economize on the battery life.

7-2 AUTO NULL

When conducting continuity tests, the contact resistance of test probes etc.

is automatically subtracted before the real resistance is displayed to obtain

more accurate reading.

This function is invalid when the contact resistance, etc. is 10

or more.

NULL symbol (

) is displayed while AUTO NULL function is working.

AUTO NULL will be cancelled when the instrument is powered off.

7-3 BACK LIGHT (Model 3007A)

Use BACK LIGHT to facilitate working at night or dimly lit situations.

When the back light switch is pressed with the function switch in any

position except "OFF", the back light continues illuminating for approx. 40

seconds and then turns off automatically.

When the back light switch is pressed again, the BACK LIGHT will turn off

even within the lighting time.

Press the back light switch while it is on, the light gets brighter. Press it

again to turn it OFF.

7-4 AUTO-POWER-OFF

The instrument automatically turns off approx. 10 minutes after the last

switch operation. To return to the normal mode, turn the function switch

off, then to the desired position.

CAUTION

Slight current is still consumed even after the instrument was powered off

by auto-power-off function. Turn the function switch to the "OFF" position

when not using the instrument.