KYORITSU 3123A User Manual

Page 5

Advertising
background image

3 −

KEW 3121A

KEW 3122A

KEW 3123A

DC Test Voltage

2500V

5000 V

5000V

10000V

Measuring Ranges

0∼2000MΩ/

1000∼100000MΩ

automatic change)

0∼5000 MΩ/

2000∼ 200000MΩ

(automatic change)

0∼5GΩ/2∼200GΩ

(automatic change)

0∼10GΩ/4∼400GΩ

(automatic change)

Accuracy Insulation

Resistance

±

5% of reading

100∼50000MΩ)

±

10% of reading

or 0.5% of scale length

ranges other than

listed above)

at 23℃ ±5℃

±

5% of reading

(200∼ 100000MΩ)

±10% of reading

or 0.5% of scale length

( ranges other than

listed above)

at 23℃ ±5℃

±

5% of reading

0.2∼100GΩ)

±

10% of reading

or 0.5% of scale length

ranges other than

listed above)

at 23℃ ±5℃

±

5% of reading

0.4∼200GΩ)

±

10% of reading

or 0.5% of scale length

ranges other than

listed above)

at 23℃ ±5℃

±

10% of reading

100∼50000MΩ)

±

20% of reading

or 1.0% of scale length

ranges other than

listed above)

at −10℃∼+40℃

±10% of reading

(200∼ 100000MΩ)

±20% of reading

or 1.0% of scale length

( ranges other than

listed above)

at −10℃ ∼+40℃

±

10% of reading

0.2∼100GΩ)

±

20% of reading

or 1.0% of scale length

ranges other than

listed above)

at −10℃∼+40℃

±

10% of reading

0.4∼200GΩ)

±

20% of reading

or 1.0% of scale length

ranges other than

listed above)

at −10℃∼+40℃

Output

Voltage

2500V ±5%

100∼50000MΩ)

5000V ±5%

(200∼ 100000MΩ)

5000V ±5%

0.2~100GΩ)

10000V ±5%

0.4~200GΩ)

Operating

Temperature & Humidity

−10℃ ∼+40℃ at 85% max. relative humidity

Storage

Temperature & Humidity

−20℃ ∼+60℃ at 90% max. relative humidity

Insulation Resistance

1000MΩ max./ 1000V between electrical circuit & housing case

Withstand Voltage

5000V AC for one minute between electrical circuit & housing case

Dimensions

200(L)×140(W)×80(D)mm

Weight

Approx. 1kg(including batteries)

Power Source

8 pcs of R6P (AA) battery or equivalent

Accessories

Hard Carrying Case:M-9158 Test Leads
(Line Probe:M-7165A, Earth
Cord:M-7224A, Guard Cord:M-7225A)

Hard Carrying Case:M-9158 Batteries, Test Leads (Line
Probe:M-7165A, Earth Cord:M-7224A, Guard Cord:M-

7225A) Pickel type prod:M-8019

Optional adaptor model 8324 is available for connection to recorder)

Advertising
This manual is related to the following products: