Specification – KYORITSU 3161A User Manual

Page 7

Advertising
background image

500V

100MΩ

2MΩ
0.1∼

50MΩ

― 5 ―

15V

20MΩ

0.05MΩ

0.005∼

2MΩ

within ±5% of indicated value

Measuring ranges other than above, 0 and ∞.

within ±10% of indicated value

within ±0.7% of scale length

0%∼+20% of nominal voltage

¡

Applicable standards
IEC 61557

(except for 15V/ 20MΩ range)

IEC 61010-1 Measurement CAT.III 300V/ CAT.II 600V

Pollution degree2, Protection class II
Location for use: altitude 2000m or less

IEC 61010-031
IEC 60529

IP40

¡

Measuring range and accuracy
<Insulation resistance range>

3.Specification

Nominal voltage
Max. effective
scale value
Mid-scale value
Accuracy in
primary effective
measuring ranges
Accuracy in
secondary effective
measuring ranges
Accuracy at 0 and ∞
Accuracy at
no-load voltage
Nominal current
Short-circuit
current
Response time

Heavy-line circular arc on the scale indicates the primary
effective measuring ranges. (guaranteed accuracy range)

1mA 0%∼+20%

Not specified

Approx. 2mA

Meter reading comes within accuracy within 3 sec.

(It may take time when measuring a capacitive load.