Communication settings, Command references – KYORITSU 8241 USB Communication Set User Manual

Page 2

Advertising
background image

Communication Settings

The communication settings for digital multimeters are fixed as follows.
   

Baud rate

: 9600 bps

   

Parity

: None

   

Stop bit

: 2 bits

   

Data length

: 8 bits

   

Handshake

: None

   

Delimiter

: <CrLf> fixed

Command References

Request to Send Real-time Data

RR command

Description

Used to acquire current measured value. The data can be
consecutively acquired with a parameter following the command.

Command

・Single data acquisition : RR,1<CrLf>
・Consecutive data acquisition : RR,2<CrLf>
・Consecutive acquisition stop : RR,0<CrLf>

Return

・Single and consecutive data acquisition
  

RR, S, MDDDDDDAUFFx<CrLf>

   

S : Status→ N=Normal  O=Over  B=LowBatt

   

M : Sign “−” or “+”

   

DDDDDD : measured value

   

A : auxiliary unit (k, m, etc.)

   

U : unit (V, A, etc.)

   

FF : measurement function (DC, AC, etc.)

   

x : Checksum

Example)
  

DC432.10mV → RR, N, +432.10mVDC7<CrLf>

Request to Send Save-mode Data

RS command

Description

Used to acquire the data from DMM save-mode memory. The
number of memory data can be specified with a parameter
following the command.

Command

RS, nn<CrLf>
  

nn : memory number (00 to 99) (KEW1061, KEW1062,

KEW1052)

Return

・When data with a specified number exists in save-mode memory
  

RS, S, MDDDDDDAUFFx<CrLf>

   

S : Status→ N=Normal O=Over B=LowBatt

   

M : Sign “−” or “+”

   

DDDDDD : measured value

   

A : auxiliary unit (k, m, etc.)

   

U : unit (V, A, etc.)

   

FF : measurement function (DC, AC, etc.)

   

x : Checksum

・When data with a specified number does NOT exist in save-mode

memory

  

RS5<CrLf>

Request to Send Logging-mode Data

RL command

Description

Used to acquire data from DMM logging-mode memory. The
number of memory data can be specified with a parameter
following the command.

Command

RL, nnnn<CrLf>
  

nnnn : memory number (0000 to 0999) (KEW1061)

  

nnnn : memory number (0000 to 9999) (KEW1062)

  

nnnn : memory number (0000 to 1599) (KEW1052)

Return

RL, S, MDDDDDDAUFFx<CrLf>
   

S : Status→ N=Normal O=Over B=LowBatt

   

M : Sign “−” or “+”

   

DDDDDD : measured value

   

A : auxiliary unit (k, m, etc.)

   

U : unit (V, A, etc.)

   

FF : measurement function (DC, AC, etc.)

   

x : Checksum

・When data with a specified number does NOT exist in logging-

mode memory

  

RLE<CrLf>

Request to Send Logging-mode Data Count

RN command

Description

Used to acquire the count of data in DMM logging-mode memory.

Command

RN<CrLf>

Return

RN, nnnnnx<CrLf>
  

nnnnn : logging-mode data count (00000 to 01000) (KEW1061)

  

nnnnn : logging-mode data count (00000 to 10000) (KEW1062)

  

nnnn : logging-mode data count (0000 to 1600) (KEW1052)

  

x : Checksum

Request to Send Interval of Logging-mode Data

RI command

Description

Used to acquire the measurement interval of data in DMM logging-
mode memory.

Command

RI<CrLf>

Return

RI, nnnnx<CrLf>
  

nnnn : interval of logging-mode data (0001 to 1800) sec

  

x : Checksum

About This Manual

1. The contents of this manual are subject to change without prior notice.

2. Extreme care has been taken in the preparation of this manual. Should, however,

any questionable descriptions or missing information be found, contact your
nearest sales representative.

3. Kyoritsu assumes no liability whatsoever for damages or injuries for which the

cause is attributable to mishandling by the user.

4. The intention of this manual is to explain the product functions and Kyoritsu makes

no warranty of any kind with regard to this manual, including but not limited to,
implied warranties of merchantability and suitability for a particular purpose.

After-sales Servicing

Should the product fail during operation or be not able to operate normally any

more, and thus need to be repaired, contact the sales representative from which you
purchased it.

9-08

92-2000A

Advertising