Table 1 Process Measurement Inputs

No. Name

Description

1, 2

Input Terminals

These terminal are used to measure current, voltage and a
contact closure for switch test.

3

P1 Pressure Port

This is the connection for the internal sensor P1

4

P2 Pressure Port

This is the connection for the internal sensor P2

5

RTD Probe Connector

This connector is where the RTD probe is plugged in.

6

Serial Interface

This is used to interface to optional external modules as
well as for RS-232 serial communications with a PC using
the LEM-232 cable(optional).

Figure 2 shows the location of the keys. Table 2 describes the function of each key.

Figure 2

Keypad

Table 2 Key Functions

No. Name

Description

1

Function Keys

These keys are used in various ways, primarily to config-
ure the calibrator

2

ON/OFF Key

This key is used to turn the calibrator on and off

3

ZERO Key

This key is used to zero pressure measurements

4

Backlight Key

This key is used to turn the backlight on and off

Note: When the calibrator is turned on by pressing the ON/OFF key, it will go through a
short startup self-check routine. During that routine, the display shows the current
firmware revision level, auto shutdown status and the ranges of the 2 internal pressure
sensors. The calibrator requires a maximum of 5 minutes warm-up to rated accuracy.
Large changes in ambient temperature may require a longer warm-up period. See
section 2.3 for instructions on zeroing the pressure sensor displays. Pressure ranges
should be zeroed each time the calibrator is started.