ERROR _ PORT?

Returns the current pressure port for percent error mode.

For example:

If you send ERROR _PORT?, it will return P1 if the pressure
port in percent error is [P1].

FUNC

Sets the display indicated in argument one to the function
indicated in argument 2.

For example:

To set the lower display to pressure mode send FUNC
LOWER,[P1].

FUNC?

Returns the current mode of all displays. For example if the
calibrator is set to [P2] ST on the upper display, [P1] on the
middle, and [P1] on the lower, FUNC? Would return:

ST_P2,P1,[P1]

HART_ON

Turns the Hart resistor on.

HART_OFF

Turns the Hart resistor off.

HART?

Returns the state of the Hart resistor.

For example:

If the Hart resistor was on HART? Would return ON.

HI_ERR

Sets the 100% point for the percent error mode calculation in the
current engineering units.

For example:

To set the 100% point to 100 psi send HI_ERR 100.