Martel Electronics MC1010 User Manual

Page 41

Advertising
background image

41

Table 12: Frequency Measurement/Source

Range

Accuracy (% of reading ± floor)

Read

2.0CPM - 600.0CPM

0.05% ± 0.1CPM

1.0Hz - 1000.0Hz

0.05% ± 0.1Hz

1.00KHz - 10.00KHz

0.05% ± 0.01KHz

Source

2.0CPM - 600.0CPM

0.05%

1.0Hz - 1000.0Hz

0.05%

1.00KHz - 10.00KHz

0.250%

Input voltage amplitude range on frequency is 1V to 20V zero based square wave only.Output amplitude is
adjustable from 1V to 20V, and is a square wave with 50% duty cycle.For output frequency, a slight negative
offset of approximately -0.1V is present to assure zero crossing.

Table 13: Resistance Measurement

Range

Accuracy (% of reading ± floor)

Ohms low

0.00

Ω - 400.0Ω

0.025% ± 0.05

Ω

Ohms high

401.0

Ω - 4000.0Ω

0.025% ± 0.5

Ω

Table 14: Resistance Source

Range

Excitation Current

Accuracy (% of reading ± floor)

Ohms low

5.0

Ω - 400.0Ω

0.1mA - 0.5mA

0.025% ± 0.1

Ω

5.0

Ω - 400.0Ω

0.5mA - 3mA

0.025% ± 0.05

Ω

Ohms high

400

Ω - 1500Ω

0.05mA - 0.8mA

0.025% ± 0.5

Ω

1500

Ω - 4000Ω

0.05mA - 0.4mA

0.025% ± 0.5

Ω

Note: Unit is compatible with smart transmitters and PLCs. Frequency response is <= 5ms.

Table 15: Thermocouple Measurement/Source

Range

Accuracy (% of reading ± floor)

Read (mV)

-10.000mV - 75.000mV

0.02% ± 10

μV

Source (mV)

-10.000mV - 75.000mV

0.02% ± 10

μV

Maximum current output in voltage ranges is 1mA with an output impedance of <= 1

Ω

Table 16: Thermocouple Read and Source (errors in °C)

Range (°C)

Accuracy (°C)

TC Type

Minimum

Maximum

CJC OFF

CJC ON

J

-210.0

-150.0

0.4

0.6

-150.0

1200.0

0.2

0.4

K

-200.0

-100.0

0.5

0.7

-100.0

600.0

0.2

0.4

600.0

1000.0

0.3

0.5

1000.0

1372.0

0.4

0.6