Event Status Enable Register (ESE)
Enables and disables bits in the ESR. Setting a bit to 1 enables the
corresponding bit in the ESR, and setting it to 0 disables the
corresponding bit. Cleared at power reset. Bit assignments for the
ESR and the ESE respectively are shown below.

15

14

13

12

11

10

9

8

0

0

0

0

0

0

0

0

7

6

5

4

3

2

1

0

PON

0

CME

EXE

DDE

QYE

0

OPC

PON

Power On. Set to 1 if power was turned on and off before
the Event Status Register was read.


CME

Command Error. Set to 1 when the calibrator receives an
invalid command. Entering an unsupported RTD type may
cause such an error.

EXE

Execution Error. Set to 1 when the calibrator runs into an
error while executing is last command. A parameter that
has too significant figures may cause such an error.

DDE

Device-dependent Error. Set to 1 when, for example, the
output of the calibrator is overloaded.

QYE

Query Error.


OPC

Operation Complete. Set to 1 when the calibrator has
finished executing all commands before the command
*OPC was entered.