Martel Electronics 235 User Manual

Page 18

Advertising
background image

step the calibrator output through the calibration points.

Once the desired results are achieved, press AUTO as indicated on the
calibrator display.




When “AUTO” is pressed, the calibrator again sets itself at the first cal
point and prompts the operator to press AUTO again to start the “As
Left” calibration procedure.

Pressing CE will take the calibrator back to the Test Complete menu.





If AUTO is pressed, the calibrator proceeds with the “As Left”
calibration.




When the test is finished, the calibrator will display “Test Complete,
Unit Passed or Failed”, and give the operator the option to “Review”
data or select “Done”.








Operation

230/235

13

MA mVOLTS
+4.000 +10.000
ADJUST UNIT
AUTO WHEN DONE

MA mVOLTS
+4.000 +10.000
AUTO TO START
CE TO ABORT

MA mVOLTS
+4.000 +10.000
TESTING CALPT 01
AS LEFT

TEST COMPLETE
UNIT PASSED
REVIEW DONE