Temperature, thermocouples (cont) – Martel Electronics MasterCAL User Manual
Page 127
Advertising
Calibrator
Specifications
117
Temperature, Thermocouples (cont)
Type Range
°
C
Measure °C
Source °C
1 Year
2 Year
1 Year
2 Year
U
-200 to 0
0.6
0.9
0.4
0.6
0 to 600
0.3
0.4
0.3
0.4
Sensor inaccuracies not included.
Accuracy with external cold junction; for internal junction add 0.2
°
C
Resolution: 0.1
°
C
Temperature Scale: ITS-90 or IPTS-68, selectable
Compensation: ITS-90 per NIST Monograph 175 for B,R,S,E,J,K,N,T; IPTS-68 per IEC 584-1 for B,R,S,E,J,K,T;
IPTS-68 per DIN 43710 for L,U.
Temperature Coefficient: 0.05
°
C/
°
C in the range -10 to 18
°
C and 28 to 50
°
C
Common Mode Error: 0.01
°
C/(Common Mode Volt)
Maximum Input Voltage: 30 V