Martel Electronics QM110T User Manual

Qm 110t, Single hole mounting thermometer, Dimensions

Advertising
background image

DIMENSIONS

All dimensions in mm (inches)

FITTING THE QM 110T

Drill a 5.5mm /

" hole in the panel.

not overtighten the nut as this may damage the meter.

7

/

32

Fit seal to the rear of the module. Fit the module to

the panel by passing its screw threaded stud and the wires through the hole, fitting the

washer and tightening the nut provided. Take care not to trap any of the connecting wires.

Do

QM110T is an easy to mount LCD digital panel meter.

To mount the module,

the user need only drill a single 5.5mm /

hole in the panel. The module is fitted to the panel by locating its screw threaded stud through the hole, fitting the

washer and tightening the nut provided. The module's connecting wires pass through the hollow stud into the target application, where they can be easily

plugged or soldered into place.

The module can measure ambient temperature via a built-in sensor and outside temperature via an

external sensor (not supplied). The module uses NTC temperature sensors to ensure high accuracy over a wide temperature range.

"

A rubber seal is included, providing splashproof protection for the unit when fitted to the meter during installation.

7

/

32

Single Hole Mounting
4 Measurement Ranges
12.5mm / 0.5" LCD Digit Height

Compatible with 10k NTC Sensors
5-Wire Connection
Splashproof

QM 110T

Single Hole Mounting Thermometer

REAR VIEW

Shown Actual Size

21.40

(0.84)

10.70

(0.42)

5.00

(0.20)

10.70

(0.42)

43.50 (1.71)

21.75 (0.85)

21.75 (0.85)

SAFETY

As the NTC temperature sensor is a passive device, no voltages should be applied to the input terminals. The user must ensure that the

incorporation of the meter into the user’s equipment conforms to the relevant sections of BS EN 61010 (Safety Requirements for Electrical Equipment for

Measuring, Control and Laboratory Use).

Stock Number

Easy Mounting Thermometer

QM 110T

Specification

Min.

Typ.

Max.

Unit

Measurement

°F range

-4

+428

°F

Resolution

Sample rate

3

sample/sec

Operating temperature range

-10

+50

°C

Supply voltage**

4

28

V d.c.

°C range

-10

+50

°C

ranges

°F range

+14

+122

°F

°C range

-20

+220

°C

0.5 (1)

°C (°F)

-20 to +140°C

1 (2)

°C (°F)

Resolution

+140 to +170°C

2 (4)

°C (°F)

+170 to +220°C

5 (10)

°C (°F)

Accuracy*

°C range

1

°C (±2 count)

°F range

2

°F (±4 count)

(Internal Sensor)

(Internal Sensor)

(External Sensor)

(Both Sensors)

(Both Sensors)

(External Sensor)

Internal Sensor

External Sensor

* Using external NTC sensor with 10k @ 25°C and a 25 to 85°C Beta value of 3977, for example:

BC Components 2322-640-73103. The module may require recalibration.

To ensure maximum accuracy, recalibrate periodically. Calibrate the module at top end of

measurement range.

** Do not apply more than 28V, as this may damage the meter.

Module thickness: 5.5mm (0.22")

°C

°C

Underrange display

Overrange display

CAL INT

CAL EXT

Washer

THERMOMETER

Panel

5 Wires - 30cm (12.0)

Nut

9.0

(0.35)

Seal

Advertising