Metex 2200D/P91 Dissolved Oxygen Controller/Analyzer User Manual

Page 17

Advertising
background image

Model 2200D/P91

Page 15

5.7 Status

5.7.1

The 2200D continuously checks the integrity of all stored data and monitors the
condition of the measuring system. If a fault is detected, the FAIL LED above the
display will turn red. The STATUS in the operation menu will provide a numerical
code, giving a possible cause and a suggested remedy.

5.7.2

The following table shows the display codes, causes and remedies:

Code

Possible Cause

Suggested Remedy

0

Normal Condition

No action required

1

Reserved

2

Temperature Sensor off Scale

Verify process. Check for open or short

connections.

3

D.O. reading off Scale

Verify process. Check for open or short

connections.

4*

Memory Loss

Call your AquaMetrix Rep. Or AquaMetrix

directly

5

Reserved

6

Sensor Fault

Check probe membrane for perforation

7

Factory Setting in force, as a

Perform procedure according to Section 9.2

result of the ESCAPE procedure

NOTE: Code 4 could be a serious failure so the alarm relay will in addition to the red illumination of the

FAIL LED for Model 2200D-X-1 and STATUS LED for Model 2200D-X-2, if DIP switch 5 Bank
S1 is OFF.

5.8 Relay A Setpoint (For Model 2200D-X-1 only)

5.8.1

As shipped from the factory, Relay A is configured to control decreasing D.O.
However, you may change the direction of control by changing the position of DIP
switch 4 Bank S1. Please refer to Section 4.8.

5.8.2

The relay setpoint may be at any place on the scale. To establish the setpoint
proceed as follows:

a) Press CALL to enter the MENU. If the PASSWORD function is enabled, enter

the password then press CALL to illuminate the “%sat/ppm” indicator. If the
PASSWORD function is not enabled, “%sat/ppm” will illuminate at the first press
of CALL.

b) Use the down arrow button to illuminate the PPM LED located to the right side

of the display.