4.1 Configuration of the input signal

- Turn the device on and wait until it completed its built-in segment test.
- Press button 2 for 1 second, in the device display appears ‚

dP’ (decimal point).

- Select the desired decimal point place by pressing button 2 respective button 3.
- Validate the selected value by pressing button 1. The display shows ‚

dP’ again.

- Press button 1 again, the display will show “

di.Lo“ (Di splay Low = low display value).

- Use button 2 and button 3 to select the desired value the device should display when a 4mA

resp. 0V input signal is attached.

- Validate the selected value by pressing button 1. The display shows “

di.Lo“ again.

- Press button 1 again, the display will show “

di.Hi“ (Display High = high display value).

- Use button 2 and button 3 to select the desired value the device should display when a 20mA

resp. 10V input signal is attached.

- Validate the selected value by pressing button 1. The display shows “

di.Hi“ again.

- Press button 1 again. The display will show “

Li“ (Limit = Measuring range limit).

- Use button 2 and button 3 to select the desired measuring range limit.

Display

Measuring range limit

Notes

off

Deactivated

Exceeding of the measuring range limit is

tolerable as of the measuring limit (p.r.t. hint).

on.Er

(on error)

Active, (displays error)

The meas. range limit is exactly bounded by

the input signal. When exceeding or

shortfalling the input signal an error message

will be displayed.

on.rG

(on range)

Active, (displays the

selected limit)

The meas. range limit is exactly bounded by

the input signal. When exceeding or short

falling the input signal the device will display

the selected lower/upper display value.

[e.g. humidity: when short falling or

exceeding, the device will display 0%

or 100%]

Hint:

When exceeding the measuring limit independently from the setting, the device will always

display an error message (“Err.1“ or “Err.2“).

The measuring limit is by 3.7 and 20.8 mA respective. 10.5 V (falling bellow 0V will not be

detected).