CJC CALIBRATION PROCEDURES

Power panel meter for at least 30 minutes.

Enter menu program, go to the

CAL

mode. Press the down

button and hold for approximately 1.5 seconds until the meter

reads

SHRT

, then release the button. Failure to hold the button

for 1.5 seconds will terminate the calibration process.

Press the down button until

CJC

appears.

Now that

CJC

appears, place the thermometer across the

terminal strip and do not remove until the temperature stabi-

lizes. The reading from the thermometer is going to be the new

CJC

temperature. Press and release the enter button. Use the

up or down button to set the value, then press the enter button

three times until V3.1 appears. Calibration is now complete.

CJC COMPENSATION