Field calibration – Pyromation 28-02700-T User Manual

Page 13

Advertising
background image

11

FIELD CALIBRATION

The thermometer is factory calibrated and does not
require calibration before use. The CAL function
allows single point calibration of the thermometer,
at 0°C (32°F) to compensate for thermocouple off-
set error. It is not necessary to perform a field cali-
bration to obtain specified meter accuracy. Use the
field calibration feature to improve thermometer/
probe accuracy or to compensate for thermocouple
drift.

1. Follow steps 1 through 4 in OPERATION sec-

tion.

2. Pack sensing end of probe in a container tightly

packed with crushed ice and filled with distilled
water. Allow temperature to stabilize.

3. Press and hold the CAL key for 5 seconds to

enter the calibration mode, the CAL annunciator
on the display starts blinking. Release the CAL
key.

4. If the measured temperature is from

10°C to

10°C (14 to 50°F), when the temperature read-
ing is stable, press the CAL key.

The CAL indicator will stop blinking and the read-
ing will be set to 0°C (32°F). The CAL indicator
will remain turned on, indicating a field calibra-
tion is active.

If “Err” is displayed, either the displayed reading is
outside the above limits or the batteries are weak.

CANCELING A FIELD CALIBRATION

1. Turn the thermometer off.

2. Hold the CAL key down while pressing the

ON/OFF key.

The field calibration is canceled and the thermom-
eter reverts to the default factory calibration. The
CAL annunciator is now turned off.