Siting Considerations

All the units are air cooled, air being pulled in through the front and
expelled from the rear. The ventilation grilles should never be
obstructed, therefore, as this will detract from the performance.
Also the fridge units should never be situated in an enclosed space
or near an extractor fan, this too will detract from the performance.

MINIMUM

MINIMUM

150 mm (6")

150 mm (6")

Setting up the unit

1. Check the voltage marked on the unit’s rating plate against that

of the mains supply. Connect the mains cable to a suitable plug
as in the following table. Note that the unit must be earthed
to ensure proper electrical safety.

Connections

230V

110/120V

Live

brown

black

Neutral

blue

white

Earth

yellow and green green

The fused plug supplied on the mains cable for use in the UK
with the RB-5A is fitted with the following fuse to protect the
cable.

Supply

fuse fitted

240V

5 Amp

For the RB-12A and RB-22A, if a fused plug is fitted then a fuse
of the following value must be fitted

Supply

fuse fitted

240V

13 Amp

The fuse in the unit protects the unit and the operator.

Note that units marked 230V on the rating plate work equally
well at 220V; units marked 120V on the rating plate work
equally well at 110V.