EcoTestr TDS High Waterproof Pocket Tester

Conditioning Before First Use:
1. Remove the transparent cap and add 20 mm of tap water or calibration standard

(approximately ½ full)

2. Recap the EcoTestr and soak* for at least 15 minutes
3. Rinse the cap after use
4. The sensor is now activated and ready for use

Measurement:
1. Remove cap and press

to turn on the tester

2. Dip sensor into the test solution
3. Stir once and let the reading stabilize. Note the measured value
4. To hold reading, press

. Screen flashes

HO

once, then displays measurement

with blinking unit (ppt) to indicate that tester is in the hold mode. Press

again to

cancel hold mode (

HC

)

5. Press

to shut the tester off

Note: Tester automatically shuts off after 8.5 minutes of non-use to conserve batteries.

TDS Factor Setting: The EcoTestr TDS High has a default factor setting of 0.71. If this is
suitable for your application, please skip this section and proceed on to calibrate.
1. Press

to turn off the tester

2. Press and hold

, then press

to turn the tester on. The display shows

dS

3. Press

to scroll to TDS factor value

4. Release

and wait for the auto confirmation

5. To abort TDS factor setting, press

to escape (

ESC

)

Calibration: The EcoTestr is factory calibrated. However, to ensure accuracy, calibration
is recommended. Note: Calibration standard sold separately.
1. If the sensor was stored dry, follow “Conditioning Before First Use” steps for

best results

2. Press

to turn on the tester

3. Dip sensor into calibration standard solution and wait for the value to stabilize
4. Press

to begin the calibration. Display shows

CAL

momentarily and blinks the

default reading

5. Press

until the blinking value matches the value of your calibration standard

at 25 ºC. Note: to set a calibration standard value that is lower than the blinking
value, continue to press

past the maximum value to continue with the lowest

adjustable value

6. Release

to accept the calibration value. After a few seconds, (

En

) is shown and

measurement is resumed

7. To abort calibration, press

to escape (

ESC

)

Maintenance: Clean the stainless steel sensors by periodically soaking them in alcohol
for 10 - 15 minutes. Rinse sensor with clean water after each use.

Useful Notes: To avoid cross contamination, rinse between samples and calibration
standards with clean water.

* The tester cap is not leak-proof. Please keep tester upright during soaking.

Eutech Order Code Oakton Order Code Description
ECOTDSTESTHIGH

35462-15

EcoTestr TDS High

ECPOUCH01

35624-45

Soft carrying case with belt-loop

EC4423000BT

3.00 ppt 442 calibration solution

00653-89

7.61 ppt 442 calibration solution, 480 ml

Changing Batteries:
1. To remove batteries, hold the tester on the left hand with the left thumb slide under

the back clip of the cover. This will release the back catch

2. Next, hold the cover with your right hand, at the same time release the front catch

with your right thumb. Slide up to remove the battery cover (Figure 1)

3. Replace old batteries with fresh ones. Note polarity as shown in battery compartment
4. No re-calibration is required after battery change. The tester’s non-volatile memory

function retains calibration settings

Your EcoTestr includes (4) pre-installed 1.5V alkaline batteries. LR44 or A76 battery types are
suitable and commonly available. Replace all (4) batteries together. Waiting too long to replace
the batteries can lead to inaccurate readings and is the most common cause of problems.

Error Messages:
1.

bA

– Batteries are weak and need replacement

2.

Er

– Calibration was attempted with a value that is <5 % of the full-scale range;

use a calibration standard with a higher value

– Use fresh calibration standard

– Sensor is dirty or has not been soaked

3.

Or

– Over Range. The measurement is above the range of the instrument

(including temperature measurement)

Reset Option: Reset option allows you to restore the calibration back to factory
default settings.
1. Press

to turn off tester

2. Press and hold

, then press

to turn the tester on. The display shows a

flashing

rS

(reset)

3. Press

again to confirm reset (

En

) or press

to escape (

ESC

) reset option

Warranty: This instrument is supplied with a warranty against manufacturing defects
for a period of one year from date of purchase.

Tester & Accessories Ordering Information:

Figure 1:

Removing battery cover

to change batteries

Battery

cover

Step 3

Slide out

the cover

LIFT

FT

Step 1

Slide in

left hand

thumb to

clear the

back

catch

LIFT

Step 2

Slide in

right hand

thumb to

clear the

front

catch