Thermo Fisher Scientific EcoScan DO 6 User Manual

Page 54

Advertising
background image

Instruction Manual

DO 6

46

7.6.2 User

Reset

This program clears all data including
calibration and other customized setup
functions and reset it to factory default.

You can skip Calibration Reset and

proceed straight to User Reset

From “rSt.5” menu,

1) Press HOLD/ENTER key three times

until meter displays the User Reset
menu “b.USR”

2) Press HOLD/ENTER key to enter the

menu.

3) Use

the

▲ and ▼ keys to activate or

deactivate the User Reset selection.

a. YES – activates User Reset

b. NO – deactivates User Reset

4) Press HOLD/ENTER key to confirm

the selection made.

5) If selection is ‘YES’, meter would

immediately return to measurement
mode. Otherwise it will return to the
main group menu. Press CAL to return to the measurement mode.

Figure 26: User Reset Selection

HOLD

ENTER

User Reset