Instruction Manual

PD 300

8.2

P1.0: Viewing previous pH calibration data

This mode lets you recall previous pH calibration data,
which helps you know when to re-calibrate your meter.
This is a “view only” mode.

From measurement mode:

1. Press

the

MODE

key to select pH measurement

mode if necessary.

2.

Press the SETUP key to enter Set up mode.

3. Press the ▲ or ▼ keys to scroll through

subgroups until you view parameter P1.0.

4. Press the ENTER key repeatedly to view previous

calibration data. See figure on right.

P1.1 = pH 1.68

P1.2 = pH 4.01

P1.3 = pH 7.00

P1.4 = pH 10.01

P1.5 = pH 12.45

5. When you have scrolled through all calibration

data, you will automatically return to the subgroup
menu. Press CAL/MEAS key to return to
measurement mode.

NOTE

If there is no previous calibration data at a particular

point, the primary display will show “ ----“.

Figure 21 – P1.0: View calibration

data for pH