Thermo Fisher Scientific Eutech 2700 Series User Manual

Page 40

Advertising
background image

36

Additional dissolved oxygen calibration notes:

Keep the membrane free from contact with solid objects.

Use with aqueous solutions only.

Do not submerse the probe past the immersion limit (see below).







See

Section 22—DO Electrode Maintenance

for additional information.

Calibration adjustment is limited to ±40% of the factory default value to

prevent erroneous calibrations.

If % calibration is attempted with a value of 10.1% to 49.9%, a

“Calibration Error” will result. The following table lists corresponding

calibration values in % saturation calibration mode:

% Saturation

(of factory default value)

Calibration Value

less than 10%

0%

10.1% to 49.9%

Calibration Error

50% to 150%

100%


Calibration of the concentration mode will only replace the previous

concentration calibration and does not affect the % saturation calibration.

To offset your % saturation reading to match another instrument, see

Section 15—Offset Calibration.

Perform daily calibration for best results. New calibration values will

automatically override the previous calibration.