Cadmium Electrode

Instruction Manual

18

TROUBLESHOOTING HINTS

Symptom

Possible Causes

Next Step

Out of Range

defective meter

check meter with

Reading

shorting strap

(see meter instruction manual)


defective electrode

check electrode operation


electrodes not

unplug electrodes

plugged in properly

and reset


reference electrode

be sure reference

not filled

electrode is filled


air bubble on

remove bubble by

membrane

redipping electrode


electrodes not

put electrodes

in solution

in solution


Noisy or Unstable

defective meter

check meter with

Readings (readings

shorting strap

continuously or
rapidly changing)

air bubble on

remove bubble by

membrane

re-dipping electrode


electrode exposed

soak electrode in

to interferences

cadmium standard


defective electrode

replace electrode


ISA not used

use recommended ISA


meter or stirrer

ground meter or

not grounded

stirrer



Drift (reading

samples and standards

allow solutions to come

slowly changing

at different

to room temperature

in one direction)

temperatures

before measurement


electrode exposed

check section

to complexing agents

entitled

Precipitation

and Complexation