3 setting temperature (set oc) sub-function, Etting temperature, Function – Thermo Fisher Scientific Alpha pH 2000 User Manual

Page 29: Setting temperature (set, C) sub-function

Advertising
background image

6.3

Setting temperature (Set

o

C) sub-function

SETUP

HOLD

SETUP

HOLD

O

C

SETUP

HOLD

O

SETUP

HOLD

SETUP

HOLD

O

C

SETUP

HOLD

O

C

atc

SETUP

HOLD

SETUP

HOLD

ENT

ENT

ENT

ENT

ENT

ENT

ENT

1

2

3

3

4

5

6

ENT

1. Select the “SET °C°F” sub-function, then press the ENT key.

2. Selecting temperature unit: press the

▲ or ▼ key to select the desired

temperature unit “°C” or “°F”. Press the ENT key to confirm your selection.

3. Enable/disable ATC mode: press the

▲ or ▼ key to enable (ATC ON)

or disable (ATC OFF) automatic temperature compensation. Press the
ENT key
to confirm your selection.

ATC disabled (ATC OFF):

For manual temperature compensation you can set the process and
calibration temperatures. This allows calibration at a temperature other
than the process temperature.

Example: setting a calibration temperature of 25°C lets you calibrate
using standard buffer solutions at 25°C, even if your process temperature
is different from 25°C.

4. Setting the process temperature: press the

▲ or ▼ key to adjust the

process temperature (upper value, setting range: –9.9 to 125°C / 15 to 257
°F). Press the ENT key to confirm process temperature.

5. Setting calibration temperature: press the

▲ or ▼ key to adjust the

calibration temperature (upper value, setting range: –9.9 to 125°C / 15 to
257 °F). Press the ENT key to confirm the calibration temperature.

ATC enabled (ATC ON):

6. Setting temperature offset: The display shows the current temperature

offset (upper value) and the temperature currently measured (lower value).
Compare the current temperature displayed on the controller to a
thermometer known to be accurate. Note down the correct temperature
value. Press the

▲ or ▼ key to adjust the lower value. The upper number

indicates the offset value. You can offset the temperature by up to
±10°C/±18°F. Press the ENT key to confirm your setting.

Continue with additional Setup procedures, or return to the Measurement
mode by pressing the

andkeys (escape) simultaneously.