Temperature compensation (TC) sub-function

This sub-function allows you to select the correct temperature compensation for optimum
operations.

SETUP

SETUP

SETUP

HOLD

HOLD

HOLD

4

4

4

SETUP

HOLD

4

ATC

°C

SETUP

HOLD

4

ATC

ENT

ENT

ENT

ENT

ENT

1

2

2

3

4

1. Select the “TC” sub-function and then press the ENT key.
2. Selecting type of temperature compensation (TC): Press the ▲ or ▼ key to select the suitable

type of temperature compensation:

− PUR TC = Pure water temperature compensation

Use pure water temperature compensation for applications in the pure water or ultra-
pure water industries.

Note: “PUR TC” should be selected while working with pure water, usually in the 0-
2 µS range.

− LIN TC= Linear temperature compensation

Use linear temperature compensation for all other applications.

Press the ENT key to confirm your selection.

− If linear temperature compensation was selected (LIN TC):

3. Selecting the temperature coefficient value for process liquid (P.TC): Press the ▲

or ▼ key to adjust the temperature coefficient value (setting range: 0 to 10 %, factory
setting: 2.10 %). Press the ENTER key to confirm the temperature coefficient value.

4. Selecting the temperature coefficient value for calibration liquid (C.TC): Press the ▲

or ▼ key to adjust the temperature coefficient value (setting range: 0 to 10 %, factory
setting: 2.10 %). Press the ENT key to confirm the temperature coefficient value.

Continue with additional Setup procedures, or return to the Measurement mode by pressing

the ▲ and ▼ key (escape) simultaneously.