Appendix 2: Simple Explanation on the Function of Hysteresis

The controller relay activates
when the set-point is reached.
In the reverse direction, it does
not de-activate when the value
reaches the set-point. Instead,
it continues to be active till the
value reaches the amount set
by the Hysteresis band.



Appendix 3: Factory Defaults

Resetting the controller to factory default settings (See program
P6.0) clears all calibration data and most other setup functions.
The following settings will remain unchanged:
1.

Measurement mode (pH or ORP)

2.

Temperature compensation mode (ATC or MTC)

3.

Temperature offset calibration value, if in ATC mode.

pH Defaults
pH input

Asymmetrical

offset 0.0

mV

slope 59.2

mV

Set point 1

pH 4.00 / LOW / Hysteresis pH
1.00

Set point 2

pH 10.00 / HIGH / Hysteresis
pH 1.00

mV Defaults
mV input

Asymmetrical

offset 0.00

mV

Set point 1

- 100 mV / LOW / Hysteresis 5
mV

Set point 2

900 mV / HIGH / Hysteresis 5
mV

Temperature Defaults
Temp. Compensation
mode

remains unchanged

MTC mode

reset to 25

o

C if in MTC mode

ATC mode

remains at last calibration if in
ATC mode

4.0

4.5

9.5

10.0

7.0

RELAY OFF

HYSTERESIS BAND

(DEFAULT = 0.5 pH)

FORWARD DIRECTION

REVERSE DIRECTION

SP1

SP2

RELAY ON

SP1 Set to LO

SP2 Set to HI