Thermo Fisher Scientific EC100GTSO05B Rhyton Body pH Combination Electrode User Manual

Page 2

Advertising
background image

www.thermofisher.com

PIN CONNECTOR WITH EUTECH ALPHA CONTROLLERS

Eutech

αlpha

pH 200

Eutech

αlpha

pH 500

Eutech

αlpha

pH 550/

560

Eutech

αlpha

pH 800

Eutech

αlpha

pH

1000

Eutech

αlpha

pH

2000P

Eutech

αlpha

pH

2000W

Electrode Model

Colour

Code

Pin No.

Pin No.

Pin No.

Pin No.

Pin No.

Pin No.

Pin No.

White

14 5 5 18 18 25 25

Black 13 6 6 19 19 26 26

Red 15 7 7 20 20 24 24

EC100GTSO05B (With
Pt 100 ATC &
potential matching
pin PMP)

Blue 4 8 8 21 21 27 27

HANDLING

Rinse electrode with distilled water before and after measuring a sample. Blot the end of the electrode
with lint-free paper to remove excess water.

Note: Never wipe the electrode to remove excess water – wiping can create static charges that interfere
with correct pH measurement.

CONDITIONING

pH electrodes are shipped with the electrodes moist. Remove the protective cap or rubber boot from the
bottom of the sensor and rinse the electrode with distilled or demonized water.

Place the electrode in a clean container containing one of the liquids i.e. 4.0 M KCl or pH 4.0 buffer or pH
7.0 buffer. Soak electrode for 20 minutes.

Note: Never condition the electrode in distilled water or demonized water – long term exposure to pure
water will damage the special glass membrane.

After conditioning the sensor, rinse the electrode with distilled or demonized water. The electrode is
ready for calibration and measurement.

STORING

Always keep pH electrode moist. Proper pH electrode storage maximizes electrode performance and
extends electrode life. It is best to store electrodes in clean containers filled with pH storage solution. Do
not store an electrode in distilled water or demonized water – this will cause ions to leach out of the glass
bulb and render your electrode useless.

CLEANING

The solution used to clean pH electrode depends on the presence of possible contaminants. Use the
guide below to choose the appropriate solution:

• For general cleaning soak the pH electrode in 0.1 M HCL or 0.1 M HNO

3

for 20 minutes

• For removing stubborn deposits and bacteria, soak the pH electrode in a 1:10 dilution of

household laundry bleach for 10 minutes.

• For removal of oil and grease, rinse the pH electrode with mild detergent or methanol

• For removal of protein deposits, soak the pH electrode in 1% pepsin in 0.1 M HCl for 5 minutes.