SENSOR STORAGE

If long-term storage of probes is required, empty electrolyte out of probe leaving only a little solution on
the membrane to keep it wet. Leaving a full volume of electrolyte in DO probe long-term without use will
deplete the probes’ anode. See section “Sensor RE-CONDITIONING ” for steps on how to open and close
probe housing.

SENSOR SPECIFICATIONS

Output at 100% saturation

PTFE membrane 24-42 mV

HDPE membrane 36-54 mV

Output at 0% saturation

PTFE membrane <1mV <1mV

HDPE membrane <1mV <1mV

Temperature Range

Max 50

°C

Min 0

°C

Accuracy

+2% when measuring temp equal cal temp.

Response Time

PTFE membrane 2 minutes to reach 90% of final reading

HDPE membrane 1 minute to reach 95% of final reading

Water Flow Rate

Min 2 inch/second across membrane

Wetted Materials

Body

Delrin

Membrane Teflon,

HDPE

Wiring

DO +

Red

DO –

Yellow

TC +

Red, White

TC –

Black