Description of controls – Time Electronics 1030 Voltage & Current Source User Manual

Page 7

Advertising
background image

7

1030 Technical Manual

P a g e

| 7

3.2. Description of Controls


1 / 2. Output Terminals
Output Voltage and Current is available on two front panel terminals which are suitable
for either wire compression or 4mm 'wander' plug insertion.


3. Polarity Switch
Normal or reverse polarity is selected by a toggle switch. The centre position is OFF
which provides an open circuit on the output terminals.

4. 6 way Position range switch

Switch Position

Switch Function

1

OFF

Voltage Ranges

2

0 – 1V (1mV resolution)

3

0 – 100mV (100uV resolution)

4

0 – 10mV (10uV resolution)

0 – 8V (10mV resolution), using external precision

1Kohm resistor (included)

Current Ranges

5

0 – 100mA (100uA resolution)

6

0 – 10mA (10uA resolution


5. 10 Turn fine adjust dial. Scaled 0-100. Linearly output from 0 – 100% of select
output range located on the top side of the 1030


6. 10 Turn dial potentiometer lock . Enables the output to be temporary locked.


7. Battery Level Indicator
Located on the top side of the 1030 this continuously monitors the battery voltage. A
minimum mark indicates when the batteries require recharging.


8. Recharge Socket
The mains recharger is a separate unit, the output of which is supplied via a flying lead
fitted with a non-reversible plug. Recharge time is between 12 and 14 hours.



CAUTION
UNDER NO CIRCUMSTANCES MUST AN ADDITIONAL VOLTAGE BE
CONNECTED IN SERIES WITH THE OUTPUT OF THE 1030 IN AN ATTEMPT TO
INCREASE THE VOLTAGE CAPABILITY AS THIS WILL CAUSE DAMAGE TO THE
OUTPUT CIRCUITRY.

Advertising