Alicat PCU Portable Calibration Unit User Manual

Page 10

Advertising
background image

10

Communication Select
The Communication Select mode is accessed by pressing the button below “Comm. RS-232” on the

Select Menu display. The screen will appear as shown in Figure 5 below.

Select

>
Unit ID

(A).....A

Baud

(19200)....19200

Data Rate......Fast

Comm.

UP DOWN RS-232

Figure 5. Communication Select Display

Unit ID – Valid unit identifiers are letters A-Z and @ (see Note below). This identifier allows the user

to assign a unique address to each device so that multiple units can be connected to a single RS-232

port on a computer. The Communication Select Mode allows you to view and/or change a unit’s unique

address. To change the unit ID address, press the “Select” button in the upper left corner of the display

until the cursor arrow is in front of the word “Unit ID”. Then, using the UP and DOWN buttons at the

bottom of the display, change the unit ID to the desired letter. Any ID change will take effect when the

Communication Select Screen is exited by pushing the MODE or Main button.
Note: When the symbol @ is selected as the unit ID, the device will go into streaming mode when

the Communication Select Mode is exited by pushing the MODE or Main button. See the RS-232

Communications chapter in this manual for information about the streaming mode.
Baud – The baud rate (bits per second) determines the rate at which data is passed back and forth

between the instrument and the computer. Both devices must send/receive at the same baud rate in

order for the devices to communicate via RS-232. The default baud rate for these devices is 19200

baud, sometimes referred to as 19.2K baud. To change the baud rate in the Communication Select

Mode, press the “Select” button in the upper left corner of the display until the cursor arrow is in front

of the word “Baud”. Then, using the UP and DOWN buttons at the bottom of the display, select the

required baud rate to match your computer or PLC. The choices are 38400, 19200, 9600, or 2400 baud.

Any baud rate change will not take effect until power to the unit is cycled.
Data Rate – Changing the Data Rate affects the rate at which the instrument dumps its data in the

streaming mode. Slow is ½ the Fast rate. The speed of the Fast rate is determined by the selected

baud rate. It is sometimes desirable to reduce the data rate if the communication speed bogs down the

computer’s processor (as is not uncommon in older laptops), or to reduce the size of data files collected

in the streaming mode. To change the data rate in the Communication Select Mode, press the “Select”

button in the upper left corner of the display until the cursor arrow is in front of the word “Data Rate”.

Then, using the UP and DOWN buttons at the bottom of the display, select either Fast or Slow.

Any data rate change will be effective immediately upon changing the value between Fast and Slow.

Advertising