BASIC PROGRAMMING

Keys

MENU – Switches to PROGRAM mode
UP ArrowScrolls forward through the parameter choices and increments numeric variables
RIGHT ArrowScrolls backward through the parameter choices and moves the active digit to the right
ENTER – Used to save programming information, advance to the next programming parameter, and

in the reset process

Modes

RUNNormal operating mode
PROGRAMUsed to program variables into the display

If your monitor was ordered with a Blancett flow meter, the two components ship from the factory calibrated as a set If the

monitor is a replacement, the turbine’s K factor has changed, or the monitor is being used with some other pulse generating

device, programming will be necessary

PROGRAMMING USING PULSE OUTPUT TURBINE FLOW METERS

Each turbine flow meter is shipped with either a K factor value or frequency data If frequency data is provided, the data must

be converted to a K factor before programming the monitor K factor information, when supplied, can usually be found on the

neck of the flow meter or stamped on the flow meter body The K factor represents the number of pulses per unit of volume

The K factor will be needed to program the monitor readout

Enter Program Mode

Change to PROGRAM mode by pressing MENU The mode indicator will change from RUN to PROGRAM

Programming

Keys

Mode

Indicator

Units Display

And

Programming Choices

Rate / Total Display

And

Programming Choices

RUN PROGRAM

RELAY1 RELAY2

Figure 2: B2800 flow monitor