Badger Meter M-Series M1000 User Manual

Installation manual, M-series® m1000

Advertising
background image

M-Series® M1000

Data Logging

MAG-UM-00677-EN-02 (July 2014)

Installation Manual

DESCRIPTION

The Data Logging kit can be ordered with this part number:

67354-010

The Data Logging feature records three types of events:

Totalizer/error events

Configuration change events

Startup events (power up or reset events)

Each type of event is recorded into three separate files stored on

internal memory.

OTEE:

N

Over time the data logging will reach the capacity of the

memory. Any new events to be recorded will overwrite

the oldest event on record.

Totalizer/Error Events

The capacity of the logging memory is 30,000 messages. The table

below defines the capacity of the memory configured for data

logging. On each interval the totalizers are recorded in addition to

any errors that have occurred from the last interval. To program the

interval, go to Miscellaneous > Datalog Period.

Interval

Totalizer / Error Events

15 min

up to 312 days

1 hr

up to 1250 days

6 hr

up to 20 years

12 hr

up to 40 years

24 hr

up to 80 years

Configuration Events

Each Configuration event identifies the parameter that was

modified and to what value the parameter was changed. Up to 40

configuration events can be recorded.

Startup Events

Each Startup event identifies the time and reason of the event. The

M1000 does not record the date and time of a power off. A total of

20 startup events can be recorded.

EXTRACTING THE EVENT FILES

All logged events can be extracted from the meter using the

supplied Flow Meter Tool software, which connects the meter to a

laptop via the supplied RS232 cable.

RS232 Link File Extraction

1. Identify / Configure the meter’s communication settings:

a. Navigate to Communications > Interface.

b. Set the interface to MODBUS.

OTEE:

N

The interface must be set to MODBUS. All other settings

can be set as desired by the operator and must match

those settings of the software tool.

c. Record or change other interface parameters

(parity and baud rate).

2. Connect the supplied RS232 cable into the RS232 connector of

the meter. Either connect the serial connector to a COM port or

connect it to the USB adapter.

3. Open the Flow Meter Tool installed on the laptop or PC. Go to

Start > All Programs > Badger Meter to open the Flow Meter Tool

application.

4. To configure the Flow Meter Tool software communication

settings:

a. Select Communication Settings.