C-1700n, Caution – Blue-White C-1700N CHEM-FEED Diaphragm Pump User Manual

Page 9

Advertising
background image

5.0

How To Operate The C-1700N

5.1

Adjusting the Pump Output- Standard models (fig. 5.1) - The C-1700N flow rate(s) can be adjusted
within a range of 5% -100% of maximum output (20:1 turndown ratio) by means of two mechanical,
cam type mechanisms. The mechanism adjusts the pump’s stroke length to an infinite number of settings
within the flow range. Because the pump’s output is reduced by increasing the pressure of the system
being injected into, the amount of suction lift, and the viscosity of the fluid being injected, the pump
must be over-sized to allow for these factors. Sizing the pump to allow adjustment within the midrange
is preferred to maintain accuracy. Consult the factory for individual pump model output curve data.

To adjust the pump’s output:

1. With the pump running, loosen the set screw.

2. Turn the adjustment knob to the desired setting.

3. Re-tighten the set screw.

C-1700N

C-1700N

Page 8

Page 9

5.3 Measuring the Pump’s Output - Volumetric Test.

This volumetric test will take into account individual installation factors such as line pressure, fluid
viscosity, suction lift, etc. This test is the most accurate for measuring the injector’s output in an individ-
ual installation.

1. Be sure the Injection Fittings and Footvalves/Strainers are clean and working properly.

2. With the injector installed under normal operating conditions, place the Footvalve/Strainer in a large
graduated cylinder.

3. Fill the graduated cylinder with the solution to be injected and run the injector until all air is removed
from the suction line and the solution enters the discharge tubing.

4. Refill the graduated cylinder, if necessary, and with the Footvalves completely submerged in the
solution, note the amount of solution in the graduated cylinder.

5. Run the injector for a measured amount of time and note the amount of fluid injected. A longer testing
time will produce more accurate results.

6.0

How to Maintain the C-1700N

6.1

Routine Inspection and Maintenance

The C-1700N requires very little maintenance. However, the pump and all accessories should be
checked regularly. This is especially important when pumping chemicals. Inspect all components for
signs of leaking, swelling, cracking, discoloration or corrosion. Replace worn or damaged components
immediately.

Cracking, crazing, discoloration and the like during the first week of operation are signs of severe
chemical attack. If this occurs, immediately remove the chemical from the pump. Determine which
parts are being attacked and replace them with parts that have been manufactured using more suitable
materials. The manufacturer does not assume responsibility for damage to the pump that has been
caused by chemical attack.

6.2

How to Clean the C-1700N

The C-1700N will require occasional cleaning, especially the Injection fittings, the Footvalves/Strainers,
and the pump head valves. The frequency will depend on the type and severity of service..

]

When changing the diaphragm, the pump head chamber and pump head cover should be wiped free of
any dirt and debris.

]

Periodically clean the injection/check valve assembly, especially when injecting fluids that calcify such
as sodium hypochlorite. These lime deposits and other build ups can clog the fitting, increase the back
pressure and interfere with the check valve operation. See section 4.3.4. Fig. 4.8.

]

Periodically clean the suction strainers. Fig.4.7

]

Periodically inspect the air vents located under the motor compartment and under the pump head. Clean
if necessary.

5.2

Adjusting the Pump Output - DELUXE Models (fig. 5.2)

In addition to the two cam type mechanism adjustments (section 5.1), the pump output of the C-1700N
deluxe unit equipped with an optional electronic cycle timer board can also be fine tuned by adjusting
the timer adjustment knob. The total-time cycle is factory preset and is not user adjustable. The on-time
cycle is adjustable from 5% to 100% of the total cycle time. Example: If the total-time cycle is 5 seconds
and the on-time cycle is adjusted for 20 percent, the pump will run for 1 second and turn off for 4
seconds (5 second total cycle). This cycle is repeated until either the cycle time is changed or the input
power is disconnected from the pump.

Note: When the input power is disconnected from the C-1700N, the unit will maintain the last
adjusted settings. When power is restored to the pump, the C-1700N will begin to pump using the
last time cycle setting.

FIG. 5.2

To adjust the On-Time :

Turn the timer adjustment knob located on the
junction box cover.

Clockwise increases the time on.

FIG. 5.1

Adjustment Knobs
(Cam Type Mechanism)

Cycle Timer
Adjustment Knob

Junction Box

MAX

MIN

1

2

3

4

5

6

7

8

9

MAX

MIN

1

2

3

4

5

6

7

8

9

ADJUST WHILE

RUNNING

LOOSEN SCREW

TURN KNOB

RE-TIGHTEN SCREW

Proper eye and skin protection must be worn when installing and
servicing the pump.

!

CAUTION

Advertising