Functions

4

XP2i Operation Manual

Resetting (Clearing) Recorded Peak Values

Peak values can only be cleared when displaying either a Peak High or Peak Low recorded pressure. Press the

(

clear

)

button for at least ½ second. Dashed

lines will briefl y appear across the display indicating that both Peak values have been cleared. Both Peak High and Peak Low values will then display the

current applied pressure. Pressing the

(

clear

)

button while either the Peak High icon (

HI

) or the Peak Low icon (

LO

) is displayed will not aff ect the zero

value. If you need to rezero the gauge, you must turn off both peak icons by pressing the

(

peak

)

button.

Zero

If you attempt to zero the gauge while applying a pressure which exceeds the Zero Limit (set in Confi gXP, defaults to 20 PSI),

the command will be ignored and

- - HI -

will be displayed.

To Zero the WT Series

Turn off peak indication by pressing the

(

peak

)

button repeatedly until the Peak High icon (

HI

) and Peak Low icon (

LO

) are off . Then press the

(

zero

)

button for at least ½ second when the gauge is vented to atmosphere.

The display will then briefl y fl ash all dashed lines (

), indicating that it has been re-zeroed. Absolute gauges will now indicate gauge pressure.

!

WARNING: This gauge can display zero pressure when connected to a source of pressure! Do not rely on the display indication before disconnecting—

it may not be indicating true pressure. Never disconnect pressure instrumentation without fi rst relieving system pressure!

To Clear the Zero Value on the WT Series

Turn off peak indication as described above, then press and hold the

(

zero

)

button until the display changes from (

) to (

).

This is especially useful for absolute gauges that have been zeroed to use for gauge pressure measurement.

clear

zero

clear

zero

zero