Entering commands – Crystal APM Advanced Pressure Module User Manual

Page 12

Advertising
background image

Communications

10

APM CPF Series Operation Manual

„„„„

„

ENTERING COMMANDS

Commands for the APM may be entered in upper or lower case. At least one space must be entered between the command and parameter. All other spaces
are optional. Almost all commands for the APM are sequential, and any overlapped commands will be indicated as such. This section will briefly explain each
of the commands and describe their general use. The descriptions will include any parameters that may be entered with the command, as well as the output
of the command.

Common Commands

X

*CLS

Clears the error queue. Also terminates all pending operations. When writing programs, use before each procedure to avoid buffer overflow.

X

*IDN?

Returns the manufacturer, model number, and firmware revision of the APM. For example,

*IDN?

will return

AMETEK, APM 0, 1.00

.

Module Commands

X

DAMP

Enables a damping filter on the readings returned by the

VAL?

command. The filter is a first-order smoothing filter with 90% damping percentage.

For example,

DAMP ON

, will turn the damping function on.

X

DAMP?

Returns the current state of the damping function. For example,

DAMP?

will return

ON

if the damping function is on.

X

FAULT?

Returns the error code number of an error that has occurred. The command may be entered when a previous command did not do what it was intended to
do. Refer to the

Error Codes table

for more information on error code numbers.

X

PRES_UNIT

Used to set the pressure unit for the indicated display. For example,

PRES_UNIT PSI

will set the pressure unit to psi. Send

CAL_STORE

to save the new

parameters. Refer to the

Parameter Units table

for a complete list of available pressure units.

X

PRES_UNIT?

Returns the current pressure unit being used.

X

VAL?

Returns the value of the current measurement on the upper and lower display. For example, if the APM is measuring 5 psi, then

VAL?

will return

5.000000E+00, PSI

.

X

ZERO_MEAS

Zeros the APM. Send

CAL_STORE

to save the new parameters.

X

ZERO_MEAS?

Returns the zero offset of the APM.