GaugeCalHP Rebuild Kit Instructions

5

PN

5138

PN

5138

Manifold Assembly

1

From the underside of the Manifold,
use a 3/32" Allen key to remove the
Retainers securing the two outside
Fittings.

2

From the topside of the Manifold,
remove the two outside Fittings.

3

Remove the six O-rings (

P N

5138

)

in the Manifold—two O-rings from
each of the three Manifold fitting
ports.

4

Rebuild the Manifold Assembly by
following the preceding steps in
reverse order, and replacing the six
O-rings (

P N

5138

).

22

Remove the Gland Assembly from
the four threaded Studs.

Rebuild the Gland Assembly using
the

procedure on page 8.

23

Once the required maintenance has
been performed on all assemblies,
reassemble the comparator by fol-
lowing steps 1-22 in reverse order.