Agilent Technologies Agilent B2201A User Manual

Page 53

Advertising
background image

Agilent B2200 User’s Guide, Edition 4

2-17

Installation

Output Connections

The 4155/4156 semiconductor parameter analyzer's Intlk connector provides the
interlock signal and a LED drive signal. If a LED is connected between pin 4 and
pin 5 (or 6) of the interlock connector, the LED lights to indicate high voltage output
when more than ±40 V is forced from an SMU in the 4155/4156.

Figure 2-6

Dimensions of LED (Agilent part number 1450-0641)

To Connect Interlock Circuit to Instrument

Before beginning the measurement, connect the interlock circuit to the instrument’s
interlock connector as follows.

For the instruments which has a BNC-Type interlock connector:

1. Get the following parts.

Agilent 16493J Interlock cable, 1 ea.

Agilent 16435A Interlock cable adapter, 1 ea.

2. Connect the 16493J interlock cable between the interlock circuit and the

16435A adapter.

3. Connect the BNC cable (furnished with the adapter) between the adapter and

the instrument’s interlock connector.

For Agilent 4155/4156/E5260/E5270:

Connect the 16493J interlock cable between the interlock circuit and the
instrument’s interlock connector. You can connect it directly without using any
adapter.

6

10

11

5

5.6

5

C

a

th

od

e(

-)

A

n

od

e(

+

)

Units: mm