Dwyer BCHP User Manual

Dwyer instruments, inc, Series bchp low pressure hand pump, Bulletin te-bchp

Advertising
background image

Bulletin TE-BCHP

The Series BCHP Calibration Test Pump is able to generate pressure
and vacuum for adjusting or calibrating pressure gauges, transmitters, or
switches. The pump is hand operated and has a pneumatic pressure
range of -28 in Hg to 870 psi (-0.95 bar to 60 bar). Dual pressure ports
allow two instruments to be connected to the pump without additional
fittings. The Series BCHP can be used in laboratories and production
areas. The fine adjustment valve on this pump ensures precise
measurements.

PRODUCT OVERVIEW

DWYER INSTRUMENTS, INC.

Phone: 219/879-8000

www.dwyer-inst.com

P.O. BOX 373 • MICHIGAN CITY, INDIANA 46360, U.S.A.

Fax: 219/872-9057

e-mail: [email protected]

SPECIFICATIONS:
Media:
Air.
Output Ranges: -28 in Hg to 870 psi (-0.95 to 60 bar).
Process Connections: 1/4˝ female BSP
Gauge Connection: 1/2˝ female BSP.
Materials: Anodized aluminum, brass, and ABS.
Weight: 8.4 lb (3.8 kg).

Specifications - Installation and Operating Instructions

Series BCHP Low Pressure Hand Pump

11/32

[8.67]

9/64

[3.61]

11/32

[8.91]

3-3/16

[81.02]

43/64

[16.88]

APPROX. 18˝

1/4˝ FEMALE BSP W/

O-RING TEST GAGE

CONNECTION

29/64

[11.54]

7/8

[22.07]

61/64

[24.26]

29/64

[11.43]

1-13/32

[35.83]

3-3/16

[81.02]

1/2˝ FEMALE BSP

W/ O-RING

REFERENCE PORT

4-27/64

[112.40]

1-25/32

[45.16]

2-3/64

[51.85]

3-1/64

[76.57]

Figure 1

1. Pressure connector

for reference instrument
1/2˝ BSP female rotating

2. Fine adjustment knob
3. Pressure relief knob

4. Change-over switch for

pressure/vacuum generation

5. Pressure connector for test

instrument 1/4˝ BSP female
rotating

6. Pump handle

Do not use tape sealant to seal the ports, as it may damage the
pump.

NOTICE

MOUNTING INSTRUCTIONS
The reference instrument is to be mounted on the top connection of the
pump. Fastening with the knurled nut by hand is sufficient. An integral O-
ring seals the reference instrument pressure port without the use of
teflon tape.

The test instrument is attached to the connection at the end of the hose.
The optional NPT adapter set includes sealing gaskets and different
adapters to best fit the instrument being tested. The tube can also be
removed to attach the test instrument directly to the pump using the
same adapter.