FUTEK VCal Sensor Verification User Manual

Page 33

Advertising
background image

Futek Advanced Sensor Technology – Futek VCal™Documentation

Rev. 1.8.0

07/26/05

©Futek Advanced Sensor Technology 2005

33

The final tab under the

Information Tabs

section of the Calibration / Verification interface is the

Test

tab. Here is where you obtain and

record your Offset readings (First and Final Unfixtured Zero Readings), and also where you will view the Uncertainty and Zero Return
values generated by your test results.

Under the Offsets section of the

Test

tab window is

where you will take your unfixtured zero readings,
both the First and the Final Zero Readings. To do
this you can either click into the text box next to the
Zero reading you wish to take and press Enter; or
you can click into the box and click on the

Un-Fixtured

button to the right of the text box. Either way your
unfixtured zero reading will appear in the text box.

Under the Results section of the

Test

tab window is

where you will come to view the Uncertainty factor
calculated for your particular test results. This is
also where you will come to view the Zero Return
value as a percentage of Full Scale.

The

Test Grid

makes up the other half of the Calibration

/ Verification interface window.

Here is where you will define which

Channel

you have

connected your Reference Sensor to, and which you
have connected your Test Sensor to.
It is also where you will define which

Units

you want to

use to measure your Load and your Output.

This is the

Test Grid

where you perform your tests and

view your outputs.

There are also buttons available here for you to

Tare

your display, take a

Gross

reading, and

Cancel

the

Current Test

.